Benito-González Tomás, Estévez-Loureiro Rodrigo, Pérez de Prado Armando, Martínez-Paz Emilse, Garrote Coloma Carmen, Fernández-Vázquez Felipe
Department of Cardiology, Hospital of León, León, Spain. Electronic correspondence:
Department of Cardiology, Hospital of León, León, Spain.
J Heart Valve Dis. 2016 Jul;25(4):508-511.
A 59-year-old male with a history of mitral valve replacement several years previously was admitted to the authors' institution with symptoms of advanced heart failure. Echocardiography showed a severe paravalvular dehiscence and surgery was discouraged due to high perioperative risk. A first transcatheter leak closure was unsuccessful despite the correct deployment of various occluding devices. A second percutaneous attempt was carried out with implantation of three Amplatzer Vascular Plug® III devices (five occluders were positioned in total within the oblong defect), leading to a mild residual leakage. Device embolism occurred twice during the procedure but was resolved using adequate catheter retrieval techniques.