Ma Xiaoli, Chen Yutong, Liu Xianjie, Ning Hong
Department of Ophthalmology, The First Hospital of China Medical University, Shenyang, Liaoning, China (mainland).
Med Sci Monit. 2016 Dec 28;22:5181-5189. doi: 10.12659/msm.898564.
BACKGROUND Identifying and assessing retinal nerve fiber layer defects are important for diagnosing and managing glaucoma. We aimed to investigate the effect of refractive correction error on retinal nerve fiber layer (RNFL) thickness measured with Spectralis spectral-domain optical coherence tomography (SD-OCT). MATERIAL AND METHODS We included 68 participants: 32 healthy (normal) and 36 glaucoma patients. RNFL thickness was measured using Spectralis SD-OCT circular scan. Measurements were made with a refractive correction of the spherical equivalent (SE), the SE+2.00D and the SE-2.00D. RESULTS Average RNFL thickness was significantly higher in the normal group (105.88±10.47 μm) than in the glaucoma group (67.67±17.27 μm, P<0.001). In the normal group, +2.00D of refractive correction error significantly affected measurements of average (P<0.001) and inferior quadrant (P=0.037) RNFL thickness. In the glaucoma group, +2.00D of refractive correction error significantly increased average (P<0.001) and individual quadrant (superior: P=0.016; temporal: P=0.004; inferior: P=0.008; nasal: P=0.003) RNFL measurements compared with those made with the proper refractive correction. However, -2.00D of refractive correction error did not significantly affect RNFL thickness measurements in either group. CONCLUSIONS Positive defocus error significantly affects RNFL thickness measurements made by the Spectralis SD-OCT. Negative defocus error did not affect RNFL measurement examined. Careful correction of refractive error is necessary to obtain accurate baseline and follow-up RNFL thickness measurements in healthy and glaucomatous eyes.
识别和评估视网膜神经纤维层缺损对于青光眼的诊断和管理至关重要。我们旨在研究屈光矫正误差对使用Spectralis光谱域光学相干断层扫描(SD-OCT)测量的视网膜神经纤维层(RNFL)厚度的影响。
我们纳入了68名参与者:32名健康(正常)人和36名青光眼患者。使用Spectralis SD-OCT环形扫描测量RNFL厚度。测量时分别采用等效球镜度(SE)、SE + 2.00D和SE - 2.00D的屈光矫正。
正常组的平均RNFL厚度(105.88±10.47μm)显著高于青光眼组(67.67±17.27μm,P<0.001)。在正常组中,+2.00D的屈光矫正误差显著影响平均(P<0.001)和下方象限(P = 0.037)RNFL厚度的测量。在青光眼组中,与采用适当屈光矫正测量的结果相比,+2.00D的屈光矫正误差显著增加了平均(P<0.001)和各个象限(上方:P = 0.016;颞侧:P = 0.004;下方:P = 0.008;鼻侧:P = 0.003)的RNFL测量值。然而,-2.00D 的屈光矫正误差在两组中均未显著影响RNFL厚度测量。
正离焦误差显著影响Spectralis SD-OCT测量所得的RNFL厚度值,但负离焦误差未影响所检测的RNFL测量值。在健康和青光眼眼中,为获得准确的基线和随访RNFL厚度测量值,必须仔细矫正屈光不正。