Marois Y, Boyer D, Guidoin R, Douville Y, Marois M, Teijeira F J, Roy P E
Laboratory of Experimental Surgery, Laval University, Québec, Canada.
Biomaterials. 1989 Aug;10(6):369-79. doi: 10.1016/0142-9612(89)90128-2.
Four chemically processed grafts implanted as arterial substitutes in dogs were evaluated as blood conduits in terms of patency rates, healing characteristics and biostability. Omniflow, Biopolymeric, Dardik-Biograft of the second generation and BIMA grafts were implanted as infrarenal aortic substitutes for 4, 24 and 48 h (short-term), 1, 2 and 4 wk (medium-term), 3 and 6 month (long-term), each type of graft being implanted for a complete series. The explanted grafts were evaluated macroscopically and processed for light and scanning electron microscopy. One Omniflow graft was occluded at 6 months and two Dardik-Biografts were thrombosed, one at 24 h and one at 1 month. All explanted grafts in the Biopolymeric series and in the BIMA were patent at the animals' death. Histological studies revealed frequent subintimal fibrosis in the Biopolymeric and Omniflow grafts. A peripheral inflammatory reaction was present in most grafts explanted. Scanning electron microscopy showed an aggravation of flow surface irregularities, after implantation of Omniflow, Biopolymeric and Dardik-Biograft, compared to virgin prostheses. These grafts presented an internal capsulae on the graft flow surface along the anastomoses. The capsulae slightly extended towards the centre of the graft after long-term implantation, (6 month) and corresponded to the pannus. Endothelial-like cells covered this pannus. The BIMA graft performed the most satisfactorily and retained its blood compatibility best, that is, the luminal surface was smooth with only minor thrombotic deposits and a thin pannus along the anastomotic lines.
将四种经过化学处理的移植物作为动脉替代物植入犬体内,从通畅率、愈合特性和生物稳定性方面对其作为血液管道进行了评估。Omniflow移植物、生物聚合物移植物、第二代Dardik生物移植物和BIMA移植物作为肾下腹主动脉替代物分别植入4、24和48小时(短期)、1、2和4周(中期)、3和6个月(长期),每种移植物均植入完整系列。对取出的移植物进行宏观评估,并进行光镜和扫描电镜处理。1个Omniflow移植物在6个月时闭塞,2个Dardik生物移植物发生血栓形成,1个在24小时时,1个在1个月时。生物聚合物系列和BIMA系列中所有取出的移植物在动物死亡时均保持通畅。组织学研究显示,生物聚合物移植物和Omniflow移植物内膜下纤维化常见。大多数取出的移植物存在外周炎症反应。扫描电镜显示,与原始假体相比,植入Omniflow移植物、生物聚合物移植物和Dardik生物移植物后,血流表面不规则性加重。这些移植物在移植物血流表面沿吻合口处有一个内部包膜。长期植入(6个月)后,包膜向移植物中心略有延伸,与血管翳相对应。内皮样细胞覆盖了这个血管翳。BIMA移植物表现最令人满意,血液相容性保持最佳,即管腔表面光滑,只有少量血栓沉积,沿吻合线有一层薄的血管翳。