锥形束计算机断层扫描的质量保证模体:系统文献综述

Quality assurance phantoms for cone beam computed tomography: a systematic literature review.

作者信息

de Oliveira Marcus V L, Wenzel Ann, Campos Paulo S F, Spin-Neto Rubens

机构信息

1 Department of Health Technology and Biology, Federal Institute of Bahia, Salvador, Brazil.

2 Institute of Health Sciences, Department of Interactive, Processes of Organs and Systems, Federal University of Bahia, Salvador, Brazil.

出版信息

Dentomaxillofac Radiol. 2017 Mar;46(3):20160329. doi: 10.1259/dmfr.20160329. Epub 2017 Feb 17.

Abstract

OBJECTIVES

To undertake a systematic review on quality assurance (QA) phantoms for CBCT imaging, including studies on the development and application of phantoms.

METHODS

The MEDLINE (PubMed) bibliographic database was searched until May 2016 for studies evaluating the development and use of phantoms in CBCT image QA. The search strategy was restricted to English language publications using the following combined terms: (Cone Beam CT) OR (Cone Beam Computed Tomography) OR (Cone-Beam Computed Tomography) OR (CBCT) AND (quality OR phantom). It was assessed which of the six image quality parameters stated by the European Commission could be evaluated with each phantom and which of them actually were.

RESULTS

The search strategy yielded 37 studies, which had developed and used (25 studies) or only used (12 studies) a phantom in CBCT image QA. According to the literature, in 7 phantoms, it is possible to evaluate 4 or more image quality parameters while in 11 phantoms, merely 1 parameter can be evaluated. Only two phantoms permit the evaluation of the six image quality parameters stated by the European Commission. The parameters, which can most often be evaluated using a phantom, are image density values, spatial resolution and geometric accuracy. The SEDENTEXCT phantom was used most frequently. In two studies, all quality parameters suggested by the European Commission were evaluated.

CONCLUSIONS

QA phantoms rarely allow all image quality parameters stated by the European Commission to be evaluated. Furthermore, alternative phantoms, which allow all image quality parameters to be evaluated in a single exposure, even for a small field of view, should be developed.

摘要

目的

对锥束计算机断层扫描(CBCT)成像的质量保证(QA)模体进行系统评价,包括模体的开发和应用研究。

方法

检索MEDLINE(PubMed)文献数据库至2016年5月,以查找评估模体在CBCT图像质量保证中的开发和使用情况的研究。检索策略仅限于使用以下组合词的英文出版物:(锥束CT)或(锥束计算机断层扫描)或(锥束计算机断层摄影)或(CBCT)与(质量或模体)。评估了欧盟委员会规定的六个图像质量参数中哪些可以用每个模体进行评估,以及哪些实际上进行了评估。

结果

检索策略产生了37项研究,这些研究在CBCT图像质量保证中开发并使用了(25项研究)或仅使用了(12项研究)模体。根据文献,在7个模体中,可以评估4个或更多图像质量参数,而在11个模体中,只能评估1个参数。只有两个模体允许评估欧盟委员会规定的六个图像质量参数。最常使用模体评估的参数是图像密度值、空间分辨率和几何精度。SEDENTEXCT模体使用最为频繁。在两项研究中,对欧盟委员会建议的所有质量参数进行了评估。

结论

QA模体很少能对欧盟委员会规定的所有图像质量参数进行评估。此外,应开发替代模体,即使对于小视野,也能在一次曝光中评估所有图像质量参数。

相似文献

8
Inhaled mannitol for cystic fibrosis.吸入用甘露醇治疗囊性纤维化。
Cochrane Database Syst Rev. 2018 Feb 9;2(2):CD008649. doi: 10.1002/14651858.CD008649.pub3.
9
Systemic treatments for metastatic cutaneous melanoma.转移性皮肤黑色素瘤的全身治疗
Cochrane Database Syst Rev. 2018 Feb 6;2(2):CD011123. doi: 10.1002/14651858.CD011123.pub2.
10
Eliciting adverse effects data from participants in clinical trials.从临床试验参与者中获取不良反应数据。
Cochrane Database Syst Rev. 2018 Jan 16;1(1):MR000039. doi: 10.1002/14651858.MR000039.pub2.

引用本文的文献

本文引用的文献

5
Dental cone beam CT: A review.牙科锥形束CT:综述
Phys Med. 2015 Dec;31(8):844-860. doi: 10.1016/j.ejmp.2015.09.004. Epub 2015 Oct 23.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验