Bernstein M A, Thomasson D M, Perman W H
Department of Medical Physics and Radiology, University of Wisconsin, Madison 53706.
Med Phys. 1989 Sep-Oct;16(5):813-7. doi: 10.1118/1.596304.
We show that for magnetic resonance (MR) images with signal-to-noise ratio (SNR) less than 2 it is advantageous to use a phase-corrected real reconstruction, rather than the more usual magnitude reconstruction. We discuss the results of the phase correction algorithm used to experimentally verify the result. We supplement the existing literature by presenting closed form expressions (in an MR context) for the probability distribution and first moments of the signal resulting from a magnitude reconstruction.
我们表明,对于信噪比(SNR)小于2的磁共振(MR)图像,使用相位校正的实部重建比更常用的幅度重建更具优势。我们讨论了用于通过实验验证该结果的相位校正算法的结果。我们通过给出幅度重建产生的信号的概率分布和一阶矩的闭式表达式(在MR背景下)来补充现有文献。