The linear penetration rate (LPR) has been used to estimate volumetric wear ever since metal-on-polyethylene bearings were first introduced. Annual volumetric wear and osteolysis are interrelated. The use of dual mobility cups has increased; however, the presence of two bearings needs paraclinical monitoring of wear to anticipate potential wear-related complications. Is LPR correlated to volumetric wear?
Ninety-eight dual mobility liner retrievals were mapped in three-dimensions (3D) with a fully automatic stripe light scanner. Martell's method was applied to 3D maps of the explants to determine how much the centres of the inner and outer bearings had shifted in the three planes, and in particular, penetration of the femoral head in the "z" axis. The total, inner and outer wear volumes were then compared to the shift in the centres of rotation (COR) and the pre-operative X-rays.
There was no relationship between wear and head penetration into the liner. There was a statistically significant relationship between the 3D centre of rotation shift and inner wear or total wear. Wear was multidirectional.
LPR is ineffective for estimating wear of dual mobility inserts, since its wear not two-dimensional. In the future, a 3D imaging method with very thin slices could be used to monitor the wear of dual mobility inserts by measuring the shift in the centres of rotation, once artifacts are overcome.
自从金属对聚乙烯关节轴承首次被引入以来,线性穿透率(LPR)就一直被用于估计体积磨损。每年的体积磨损和骨溶解是相互关联的。双活动臼杯的使用有所增加;然而,由于存在两个轴承,需要进行临床旁监测磨损情况,以预测潜在的与磨损相关的并发症。LPR与体积磨损相关吗?
使用全自动条纹光扫描仪对98个双活动衬垫取出物进行三维(3D)映射。将马特尔方法应用于取出物的3D地图,以确定内、外轴承中心在三个平面上的移动情况,特别是股骨头在“z”轴上的穿透情况。然后将总磨损量、内部磨损量和外部磨损量与旋转中心(COR)的移动以及术前X光片进行比较。
磨损与股骨头穿透衬垫之间没有关系。3D旋转中心移动与内部磨损或总磨损之间存在统计学上的显著关系。磨损是多方向的。
LPR对于估计双活动植入物的磨损无效,因为其磨损不是二维的。未来,一旦克服伪影,可使用具有非常薄切片的3D成像方法通过测量旋转中心的移动来监测双活动植入物的磨损情况。