使用较低碘浓度的儿科计算机断层扫描心脏血管造影中的主观和客观图像差异。

Subjective and objective image differences in pediatric computed tomography cardiac angiography using lower iodine concentration.

作者信息

Hwang Jae-Yeon, Choo Ki Seok, Choi Yoon Young, Kim Jin Hyeok, Ryu Hwaseong, Han Junhee, Kim Yong-Woo, Jeon Ung Bae, Nam Kyung Jin

机构信息

Department of Radiology, Pusan National University Yangsan Hospital, Geumo-ro 20, Mulgeum-eup, Yangsan-si, Gyeongsangnam-do, 50612, Republic of Korea.

Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Geumo-ro 20, Mulgeum-eup, Yangsan-si, Gyeongsangnam-do, 50612, Republic of Korea.

出版信息

Pediatr Radiol. 2017 May;47(6):701-709. doi: 10.1007/s00247-017-3795-z. Epub 2017 Feb 24.

Abstract

BACKGROUND

Several recent studies showed the optimal contrast enhancement with a low-concentration and iso-osmolar contrast media in both adult and pediatric patients. However, low contrast media concentrations are not routinely used due to concerns of suboptimal enhancement of cardiac structures and small vessels.

OBJECTIVE

To evaluate the feasibility of using iso-osmolar contrast media containing a low iodine dose for CT cardiac angiography at 80 kilovolts (kVp) in neonates and infants.

MATERIALS AND METHODS

The iodixanol 270 group consisted of 79 CT scans and the iopromide 370 group of 62 CT scans in patients ≤1 year old. Objective measurement of the contrast enhancement was analyzed and contrast-to-noise ratios of the ascending aorta and left ventricle were calculated. Regarding subjective measurement, a four-point scale system was devised to evaluate degrees of contrast enhancement, image noise, motion artifact and overall image quality of each image set. Reader performance for correctly differentiating iodixanol 270 and iopromide 370 by visual assessment was evaluated.

RESULTS

Group objective and subjective measurements were nonsignificantly different. Overall sensitivity, specificity and diagnostic accuracy for correctly differentiating iodixanol 270 and iopromide 370 by visual assessment were 42.8%, 59%, and 50%, respectively.

CONCLUSION

The application of iodixanol 270 achieved optimal enhancement for performing pediatric cardiac CT angiography at 80 kVp in neonates and infants. Objective measurements of contrast enhancement and subjective image quality assessments were not statistically different in the iodixanol 270 and iopromide 370 groups.

摘要

背景

最近的几项研究表明,低浓度等渗造影剂在成人和儿童患者中均可实现最佳的对比增强效果。然而,由于担心心脏结构和小血管的增强效果欠佳,低浓度造影剂并未常规使用。

目的

评估在80千伏峰值电压(kVp)下,使用低碘剂量的等渗造影剂进行新生儿和婴儿CT心脏血管造影的可行性。

材料与方法

碘克沙醇270组包括79例≤1岁患者的CT扫描,碘普罗胺370组包括62例≤1岁患者的CT扫描。分析对比增强的客观测量结果,并计算升主动脉和左心室的对比噪声比。对于主观测量,设计了一个四点量表系统来评估每组图像的对比增强程度、图像噪声、运动伪影和整体图像质量。评估阅片者通过视觉评估正确区分碘克沙醇270和碘普罗胺370的表现。

结果

两组的客观和主观测量结果无显著差异。通过视觉评估正确区分碘克沙醇270和碘普罗胺370的总体敏感性、特异性和诊断准确性分别为42.8%、59%和50%。

结论

碘克沙醇270的应用在80 kVp下对新生儿和婴儿进行儿科心脏CT血管造影时实现了最佳增强效果。碘克沙醇270组和碘普罗胺370组在对比增强的客观测量和主观图像质量评估方面无统计学差异。