Kruve Anneli
Institute of Chemistry, University of Tartu, Ravila 14a, Tartu, 50411, Estonia.
J Mass Spectrom. 2016 Aug;51(8):596-601. doi: 10.1002/jms.3790.
Electrospray ionization (ESI) efficiency is known to be affected by the properties of the analytes, source design and source parameters. In this study, the ionization efficiency of 17 acidic compounds at various conditions in ESI negative ion mode was evaluated. Namely, the influence of organic solvent content in the mobile phase, ionization source parameters, ionization source geometry and functionality (conventional ESI, ESI with thermal focussing and with additional internal nebulizer gas) was studied. It was observed that the ionization efficiency in thermal focussing ESI is only marginally affected by the organic solvent composition, while for conventional ESI and ESI with internal nebulizer gas, the ionization efficiency increases significantly with increasing organic modifier content. For all ionization sources and mobile phase compositions, the ionization efficiency values between different setups showed good correlation. Copyright © 2016 John Wiley & Sons, Ltd.
已知电喷雾电离(ESI)效率受分析物性质、源设计和源参数的影响。在本研究中,评估了17种酸性化合物在ESI负离子模式下各种条件下的电离效率。具体而言,研究了流动相中有机溶剂含量、电离源参数、电离源几何形状和功能(传统ESI、具有热聚焦功能的ESI和带有额外内部雾化气的ESI)的影响。观察到,热聚焦ESI中的电离效率仅受有机溶剂组成的轻微影响,而对于传统ESI和带有内部雾化气的ESI,电离效率随有机改性剂含量的增加而显著提高。对于所有电离源和流动相组成,不同设置之间的电离效率值显示出良好的相关性。版权所有© 2016约翰威立父子有限公司。