Penton Z
Varian Instrument Group, Walnut Creek, CA 94598.
Clin Chem. 1987 Nov;33(11):2094-5.
Several 0.53 mm (i.d.) fused-silica open tubular columns were evaluated for the gas-chromatographic determination of ethanol and other volatiles in blood by both headspace and liquid injection. These columns offer the advantages of fused-silica technology without requiring expensive modification of the instrument for capillary. Methyl silicone columns of various lengths and film thicknesses were examined and also a polyethylene glycol column. The merits of each column are discussed and relative retention times are given.
对几根内径为0.53毫米的熔融石英开管柱进行了评估,以通过顶空进样和液体进样气相色谱法测定血液中的乙醇和其他挥发性物质。这些色谱柱具有熔融石英技术的优势,无需对仪器进行昂贵的毛细管改装。研究了不同长度和膜厚的甲基硅氧烷柱以及一根聚乙二醇柱。讨论了每根色谱柱的优点并给出了相对保留时间。