智能手机光电容积脉搏波法与 Polar H7 胸带和心电图心率变异性记录的比较。

Comparison of Heart-Rate-Variability Recording With Smartphone Photoplethysmography, Polar H7 Chest Strap, and Electrocardiography.

出版信息

Int J Sports Physiol Perform. 2017 Nov 1;12(10):1324-1328. doi: 10.1123/ijspp.2016-0668. Epub 2017 Dec 22.

Abstract

PURPOSE

To establish the validity of smartphone photoplethysmography (PPG) and heart-rate sensor in the measurement of heart-rate variability (HRV).

METHODS

29 healthy subjects were measured at rest during 5 min of guided breathing and normal breathing using smartphone PPG, a heart-rate chest strap, and electrocardiography (ECG). The root mean sum of the squared differences between R-R intervals (rMSSD) was determined from each device.

RESULTS

Compared to ECG, the technical error of estimate (TEE) was acceptable for all conditions (average TEE CV% [90% CI] = 6.35 [5.13; 8.5]). When assessed as a standardized difference, all differences were deemed "trivial" (average standard difference [90% CI] = 0.10 [0.08; 0.13]). Both PPG- and heart-rate-sensor-derived measures had almost perfect correlations with ECG (R = 1.00 [0.99; 1.00]).

CONCLUSION

Both PPG and heart-rate sensors provide an acceptable agreement for the measurement of rMSSD when compared with ECG. Smartphone PPG technology may be a preferred method of HRV data collection for athletes due to its practicality and ease of use in the field.

摘要

目的

验证智能手机光电体积描记法(PPG)和心率传感器在心率变异性(HRV)测量中的有效性。

方法

使用智能手机 PPG、心率胸带和心电图(ECG),对 29 名健康受试者在 5 分钟的指导呼吸和正常呼吸期间的心率进行测量。从每个设备中确定 RR 间期均方根差的均值(rMSSD)。

结果

与 ECG 相比,所有条件下的技术估计误差(TEE)均为可接受(平均 TEE CV%[90%CI]=6.35[5.13;8.5])。当作为标准化差异进行评估时,所有差异均被认为“微不足道”(平均标准差异[90%CI]=0.10[0.08;0.13])。PPG 和心率传感器衍生的测量值与 ECG 具有几乎完美的相关性(R=1.00[0.99;1.00])。

结论

与 ECG 相比,PPG 和心率传感器在测量 rMSSD 时具有可接受的一致性。由于智能手机 PPG 技术在现场的实用性和易用性,它可能成为运动员 HRV 数据采集的首选方法。