Eames W B, Sieweke J C, Wallace S W, Rogers L B
J Prosthet Dent. 1979 Mar;41(3):304-7. doi: 10.1016/0022-3913(79)90013-1.
Impression trays were fabricated providing 2, 4, and 6 mm spaces to determine the stability and accuracy of nine elastomeric impression materials on a simulated full crown preparation steel die. The interface space of 2 mm produced the most accurate impressions for all of the materials tested. All impression materials except one fell within the revised American Dental Association Specifications. The clinical-type tests, using castings on dies poured from these materials, corroborated the acceptance of those materials and techniques exhibiting the least dimensional change.