Diagnostic value of optimised real-time sonoelastography in the assessment of liver fibrosis in chronic hepatitis B and C.

作者信息

Kalita Katarzyna, Filipczak Krzysztof, Bieńkiewicz Małgorzata, Deroń Wojciech, Deroń Zbigniew, Piekarska Anna, Płachcińska Anna, Kuśmierek Jacek

机构信息

Department of Nuclear Medicine, Ultrasound Section, Medical University of Lodz, Lodz, Poland.

Department of Quality Control and Radiological Protection, Medical University of Lodz, Lodz, Poland.

出版信息

Prz Gastroenterol. 2017;12(1):28-33. doi: 10.5114/pg.2016.61475. Epub 2016 Jul 22.

Abstract

AIM

To optimise the method of real-time elastography (RTE) in the assessment of liver fibrosis using an in-house prepared method for elastogram analysis, as well as a semiquantitative analysis based on newly introduced parameters.

MATERIAL AND METHODS

Sonoelastography was performed in 94 patients with various degrees of liver fibrosis and also in 25 healthy volunteers. As a reference method for diagnostic efficacy of sonoelastography-based parameters used for the assessment of fibrosis degree in patients with chronic B and C hepatitis, a liver biopsy was used. Patient's elastograms were analysed using in-house prepared software, Pixel Count, calculating two semiquantitative parameters: mean stiffness fraction (MSF%) and intrinsic stiffness ratio (ISR).

RESULTS

Statistically significant differences between distributions of the above presented parameters for different degrees of liver fibrosis were revealed. Indices of diagnostic efficacy for detection of significant liver fibrosis ( ≥ 2) using MSF% amounted to: sensitivity - 76%, specificity - 87% and ISR: 81% and 87%, respectively. Sensitivity of both parameters in detection of cirrhosis ( = 4) was equal to 88% and specificity amounted to: for MSF% - 84% and ISR - 86%. Interobserver reproducibility determined for both of the above parameters was high, intraclass correlation coefficients (ICC) were 0.91 for MSF% and 0.93 for ISR.

CONCLUSIONS

Real-time elastography applied in this study, using in-house prepared Pixel Count software, provided good reproducibility and diagnostic efficacy, especially specificity, in the assessment of liver fibrosis degree.

摘要

目的

采用自行编制的弹性图分析方法以及基于新引入参数的半定量分析方法,优化实时弹性成像(RTE)评估肝纤维化的方法。

材料与方法

对94例不同程度肝纤维化患者及25名健康志愿者进行超声弹性成像检查。采用肝活检作为参考方法,用于评估慢性乙型和丙型肝炎患者纤维化程度的基于超声弹性成像参数的诊断效能。使用自行编制的软件Pixel Count分析患者的弹性图,计算两个半定量参数:平均硬度分数(MSF%)和固有硬度比(ISR)。

结果

不同程度肝纤维化上述参数分布之间存在统计学显著差异。使用MSF%检测显著肝纤维化(≥2)的诊断效能指标为:灵敏度-76%,特异度-87%;ISR的灵敏度和特异度分别为81%和87%。两个参数检测肝硬化(=4)的灵敏度均为88%,特异度分别为:MSF%-84%,ISR-86%。上述两个参数的观察者间重复性较高,组内相关系数(ICC):MSF%为0.91,ISR为0.93。

结论

本研究应用自行编制的Pixel Count软件进行实时弹性成像,在评估肝纤维化程度方面具有良好的重复性和诊断效能,尤其是特异度。