Predicting ESI/MS Signal Change for Anions in Different Solvents.

机构信息

University of Tartu , Institute of Chemistry, Ravila 14a, Tartu 50411, Estonia.

Technion - Israel Institute of Technology , Schulich Faculty of Chemistry, Technion City, Haifa 3200008, Israel.

出版信息

Anal Chem. 2017 May 2;89(9):5079-5086. doi: 10.1021/acs.analchem.7b00595. Epub 2017 Apr 17.

Abstract

LC/ESI/MS is a technique widely used for qualitative and quantitative analysis in various fields. However, quantification is currently possible only for compounds for which the standard substances are available, as the ionization efficiency of different compounds in ESI source differs by orders of magnitude. In this paper we present an approach for quantitative LC/ESI/MS analysis without standard substances. This approach relies on accurately predicting the ionization efficiencies in ESI source based on a model, which uses physicochemical parameters of analytes. Furthermore, the model has been made transferable between different mobile phases and instrument setups by using a suitable set of calibration compounds. This approach has been validated both in flow injection and chromatographic mode with gradient elution.

摘要

LC/ESI/MS 是一种广泛应用于各个领域的定性和定量分析技术。然而,目前只有对于有标准物质的化合物才能进行定量分析,因为在 ESI 源中不同化合物的离子化效率差异很大。本文提出了一种无需标准物质的定量 LC/ESI/MS 分析方法。该方法依赖于基于模型准确预测 ESI 源中的离子化效率,该模型使用分析物的物理化学参数。此外,通过使用合适的校准化合物集,该模型已经可以在不同的流动相和仪器设置之间转移。该方法在流动注射和梯度洗脱的色谱模式下都进行了验证。