Suppr超能文献

一种用于即时检测符合指南的心肌肌钙蛋白T检测的新型免疫化学平台:原理验证

A new immunochemistry platform for a guideline-compliant cardiac troponin T assay at the point of care: proof of principle.

作者信息

Lopez-Calle Eloisa, Espindola Pamela, Spinke Juergen, Lutz Sascha, Nichtl Alfons, Tgetgel Armin, Herbert Nicolas, Marcinowski Moritz, Klepp Juergen, Fischer Thomas, Brueckner Thorsten, Boehm Christoph, Keller Thomas

机构信息

.

出版信息

Clin Chem Lab Med. 2017 Oct 26;55(11):1798-1804. doi: 10.1515/cclm-2016-1078.

Abstract

BACKGROUND

A multitude of troponin assays for the point-of-care (POC) have been developed showing a lack of analytical sensitivity and precision. We present a new platform solution for the high-sensitivity detection of cardiac troponin T (cTnT) in a 30 μL whole blood sample with a turnaround time of 11 min.

METHODS

The immunoassay was completely run in a ready-to-use plastic disposable, a centrifugal microfluidic disc with fully integrated reagents. After the sample application, the assay was automatically processed by separating the cellular blood components via centrifugation, followed by incubation of a defined volume from the generated plasma with the immunoreagents. The fluorescence in the signal zone of a membrane was measured after its washing for the cTnT quantitation.

RESULTS

A calibration curve, measured in whole blood samples spiked with native human cTnT, was generated covering a range up to a concentration of approximately 8300 ng/L. The lower detection limit was determined to be 3.0 ng/L. At a concentration of 14 ng/L, the 99th percentile value from the high-sensitivity cardiac troponin T (hs-cTnT) assay in the Elecsys® system, the imprecision (CV) was 3.8%. A CV profile indicated that the functional sensitivity for a CV <10% was 6.8 ng/L. The assay did not show any significant cross-reaction with human skeletal troponin T. We observed an excellent correlation with the hs-TnT Elecsys® assay for 49 clinical plasma samples (r=0.9744).

CONCLUSIONS

The described technology shows that an analytical performance for a highly sensitive determination of cTnT can be achieved in a POC setting.

摘要

背景

已经开发出多种用于即时检测(POC)的肌钙蛋白检测方法,但这些方法显示出分析灵敏度和精密度不足。我们提出了一种新的平台解决方案,可在30μL全血样本中高灵敏度检测心肌肌钙蛋白T(cTnT),周转时间为11分钟。

方法

免疫测定完全在即用型一次性塑料离心微流控盘上进行,该盘带有完全集成的试剂。加样后,通过离心分离血细胞成分自动处理该检测,然后将产生的血浆中确定体积的样本与免疫试剂孵育。洗涤膜的信号区后测量荧光以进行cTnT定量。

结果

在添加天然人cTnT的全血样本中测量得到校准曲线,其浓度范围高达约8300 ng/L。测定的下限为3.0 ng/L。在浓度为14 ng/L时,即Elecsys®系统中高灵敏度心肌肌钙蛋白T(hs-cTnT)检测的第99百分位数值,不精密度(CV)为3.8%。CV曲线表明,CV<10%时的功能灵敏度为6.8 ng/L。该检测与人骨骼肌肌钙蛋白T没有任何显著的交叉反应。对于49份临床血浆样本,我们观察到与Elecsys® hs-TnT检测具有极好的相关性(r=0.9744)。

结论

所描述的技术表明,在即时检测环境中可以实现对cTnT的高灵敏度分析性能。