Elecsys® Troponin T Gen 5 STAT 检测试剂分析性能评估。

Analytical performance evaluation of the Elecsys® Troponin T Gen 5 STAT assay.

机构信息

Department of Pathology, UC San Diego Health, San Diego, CA, United States of America.

Department of Emergency Medicine, Thomas Jefferson University, Philadelphia, PA, United States of America.

出版信息

Clin Chim Acta. 2019 Aug;495:522-528. doi: 10.1016/j.cca.2019.05.026. Epub 2019 May 30.

Abstract

BACKGROUND

We report the analytical performance of the Elecsys® Troponin T Gen 5 STAT (TnT Gen 5 STAT; Roche Diagnostics) assay.

METHODS

Measuring limits/ranges were determined in lithium-heparin plasma samples per Clinical and Laboratory Standards Institute (CLSI) EP17-A2. Precision was evaluated per CLSI EP05-A2 using lithium-heparin plasma/quality control samples on cobas e 411/cobas e 601 analyzers; two duplicated runs per day for 21 days (n = 84). Cross-reactivity with other troponin forms and interference from endogenous substances/drugs was tested; recovery criterion for no cross-reactivity was within ±10%.

RESULTS

Coefficients of variation (CV) for repeatability/intermediate precision were 0.7-5.6%/1.4-10.3% (cobas e 411; mean cardiac troponin T [cTnT]: 7.3-9341 ng/L) and 0.7-3.0%/1.5-6.4% (cobas e 601; mean cTnT: 7.4-9455 ng/L). There was no cross-reactivity with skeletal muscle troponin T (≤ 10,000 ng/L), skeletal muscle troponin I (≤ 100,000 ng/L), cardiac troponin I (≤ 10,000 ng/L), or human troponin C (≤ 80,000 ng/L). No interference was observed with biotin (≤ 20 ng/mL) or 34 drugs.

CONCLUSION

The TnT Gen 5 STAT assay demonstrated a CV of <10% at the 99th percentile upper reference limit, meeting precision requirements (Fourth Universal Definition of Myocardial Infarction) for high-sensitivity troponin assays.

摘要

背景

我们报告 Elecsys® Troponin T Gen 5 STAT(TnT Gen 5 STAT;罗氏诊断)分析性能。

方法

根据临床和实验室标准协会(CLSI)EP17-A2 测定锂肝素血浆样本中的测定下限/范围。使用 cobas e 411/cobas e 601 分析仪上的锂肝素血浆/质控样本,根据 CLSI EP05-A2 评估精密度;每天重复 2 次,共 21 天(n=84)。测试与其他肌钙蛋白形式的交叉反应性和内源性物质/药物的干扰;无交叉反应性的恢复标准为±10%以内。

结果

重复性/中间精密度的变异系数(CV)为 0.7-5.6%/1.4-10.3%(cobas e 411;平均心肌肌钙蛋白 T [cTnT]:7.3-9341ng/L)和 0.7-3.0%/1.5-6.4%(cobas e 601;平均 cTnT:7.4-9455ng/L)。与骨骼肌肌钙蛋白 T(≤10000ng/L)、骨骼肌肌钙蛋白 I(≤100000ng/L)、心肌肌钙蛋白 I(≤10000ng/L)或人肌钙蛋白 C(≤80000ng/L)无交叉反应。未观察到生物素(≤20ng/mL)或 34 种药物的干扰。

结论

TnT Gen 5 STAT 分析在第 99 个百分位上限的 90%时具有<10%的 CV,满足高敏肌钙蛋白分析的精密度要求(心肌梗死的第四通用定义)。