硅胶净化后氢化物发生-原子吸收光谱法测定生物样品中的无机锡。

Determination of inorganic tin in biological samples by hydride generation-atomic absorption spectrometry after silica gel cleanup.

作者信息

Tsuda T, Wada M, Aoki S, Matsui Y

机构信息

Shiga Prefectural Institute of Public Health and Environmental Science, Japan.

出版信息

J Assoc Off Anal Chem. 1988 Mar-Apr;71(2):373-4.

PMID:2838449
Abstract

A method is described for the determination of inorganic tin in biological samples by hydride generation-atomic absorption spectrometry (HG-AAS). A sample is extracted with ethyl acetate after addition of HCl and NaCl. The concentrated extract is passed through a silica gel column. The column is washed with ethanol, water, and 0.2N HCl successively, and then inorganic tin is eluted with 2N HCl and measured by HG-AAS. Recoveries from fish muscle spiked with 0.1 micrograms/g Sn4+ are 78.9 +/- 4.2% (average +/- standard deviation, n = 5). The detection limit is 0.01 micrograms/g as Sn.

摘要

描述了一种采用氢化物发生-原子吸收光谱法(HG-AAS)测定生物样品中无机锡的方法。在加入HCl和NaCl后,用乙酸乙酯萃取样品。将浓缩提取物通过硅胶柱。依次用乙醇、水和0.2N HCl洗涤柱子,然后用2N HCl洗脱无机锡,并用HG-AAS进行测定。在添加了0.1微克/克Sn4+的鱼肉中的回收率为78.9±4.2%(平均值±标准偏差,n = 5)。以Sn计的检测限为0.01微克/克。