Abstract

PURPOSE

Reduced bone strength is associated with a loss of bone mass, usually evaluated by dual-energy X-ray absorptiometry, although it is known that the bone microstructure also affects the bone strength. Here, a method is proposed to measure (in laboratory) the bone volume-to-total volume ratio by single-sided NMR scanners, which is related to the microstructure of the trabecular bone.

METHODS

Three single-sided scanners were used on animal bone samples. These low-field, mobile, low-cost devices are able to detect the NMR signal, regardless of the sample sizes, without the use of ionizing radiations, with the further advantage of signal localization offered by their intrinsic magnetic field gradients.

RESULTS

The performance of the different single-sided scanners have been discussed. The results have been compared with bone volume-to-total volume ratio by micro CT and MRI, obtaining consistent values.

CONCLUSIONS

Our results demonstrate the feasibility of the method for laboratory analyses, which are useful for measurements like porosity on bone specimens. This can be considered as the first step to develop an NMR method based on the use of a mobile single-sided device, for the diagnosis of osteoporosis, through the acquisition of the signal from the appendicular skeleton, allowing for low-cost, wide screening campaigns. Magn Reson Med 79:501-510, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

摘要

目的

骨强度的降低与骨量的减少有关,通常通过双能 X 射线吸收法来评估,尽管众所周知,骨微结构也会影响骨强度。在这里,提出了一种通过单边 NMR 扫描仪测量(实验室中)骨体积与总体积比的方法,该方法与小梁骨的微观结构有关。

方法

在动物骨样本上使用了三台单边扫描仪。这些低场、移动、低成本的设备能够检测到 NMR 信号,而不管样本大小如何,无需使用电离辐射,并且由于其固有磁场梯度提供了信号定位的进一步优势。

结果

讨论了不同单边扫描仪的性能。将结果与微 CT 和 MRI 的骨体积与总体积比进行了比较,得到了一致的值。

结论

我们的结果证明了该方法在实验室分析中的可行性,这对于骨标本的孔隙率等测量很有用。这可以被认为是开发一种基于使用移动单边设备的 NMR 方法的第一步,用于通过从附肢骨骼获取信号来诊断骨质疏松症,从而实现低成本、广泛的筛查活动。磁共振医学 79:501-510, 2018. © 2017 国际磁共振学会。