Assessment of Calcaneal Spongy Bone Magnetic Resonance Characteristics in Women: A Comparison between Measures Obtained at 0.3 T, 1.5 T, and 3.0 T.

作者信息

Capuani Silvia, Maiuro Alessandra, Giampà Emiliano, Montuori Marco, Varrucciu Viviana, Hagberg Gisela E, Vinicola Vincenzo, Colonna Sergio

机构信息

CNR-ISC c/o Physics Department, "Sapienza" University of Rome, P.zle Aldo Moro 5, 00185 Rome, Italy.

Neuroimaging Laboratory, Santa Lucia Foundation, IRCCS Rome, Via Ardeatina 309, 00179 Rome, Italy.

出版信息

Diagnostics (Basel). 2024 May 18;14(10):1050. doi: 10.3390/diagnostics14101050.

Abstract

BACKGROUND

There is a growing interest in bone tissue MRI and an even greater interest in using low-cost MR scanners. However, the characteristics of bone MRI remain to be fully defined, especially at low field strength. This study aimed to characterize the signal-to-noise ratio (SNR), T, and T* in spongy bone at 0.3 T, 1.5 T, and 3.0 T. Furthermore, relaxation times were characterized as a function of bone-marrow lipid/water ratio content and trabecular bone density.

METHODS

Thirty-two women in total underwent an MR-imaging investigation of the calcaneus at 0.3 T, 1.5 T, and 3.0 T. MR-spectroscopy was performed at 3.0 T to assess the fat/water ratio. SNR, T, and T* were quantified in distinct calcaneal regions (ST, TC, and CC). ANOVA and Pearson correlation statistics were used.

RESULTS

SNR increase depends on the magnetic field strength, acquisition sequence, and calcaneal location. T* was different at 3.0 T and 1.5 T in ST, TC, and CC. Relaxation times decrease as much as the magnetic field strength increases. The significant linear correlation between relaxation times and fat/water found in healthy young is lost in osteoporotic subjects.

CONCLUSION

The results have implications for the possible use of relaxation vs. lipid/water marrow content for bone quality assessment and the development of quantitative MRI diagnostics at low field strength.

摘要

背景

人们对骨组织磁共振成像(MRI)的兴趣日益浓厚,对使用低成本MR扫描仪的兴趣更是与日俱增。然而,骨MRI的特性仍有待全面界定,尤其是在低场强情况下。本研究旨在对0.3T、1.5T和3.0T场强下松质骨的信噪比(SNR)、T1和T2*进行特征描述。此外,还将弛豫时间描述为骨髓脂质/水比例含量和小梁骨密度的函数。

方法

共有32名女性在0.3T、1.5T和3.0T场强下对跟骨进行了MR成像检查。在3.0T场强下进行磁共振波谱分析以评估脂肪/水比例。在跟骨的不同区域(ST、TC和CC)对SNR、T1和T2*进行了量化。使用了方差分析和Pearson相关统计方法。

结果

SNR的增加取决于磁场强度、采集序列和跟骨位置。在ST、TC和CC区域,3.0T和1.5T场强下的T2*有所不同。弛豫时间随磁场强度增加而降低。在健康年轻人中发现的弛豫时间与脂肪/水之间的显著线性相关性在骨质疏松患者中消失。

结论

这些结果对于利用弛豫与脂质/水骨髓含量进行骨质量评估以及低场强下定量MRI诊断的发展可能具有重要意义。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验