基于剪切波弹性成像(SWE)的皮肤弹性定量评估的可重复性分析。

Reproducibility analysis on shear wave elastography (SWE)-based quantitative assessment for skin elasticity.

作者信息

Sun Yang, Ma Chuan, Liang XiaoLong, Wang Run, Fu Ying, Wang ShuMin, Cui LiGang, Zhang ChunLei

机构信息

Department of Ultrasound Department of Dermatology, Peking University Third Hospital, Beijing, P.R. China.

出版信息

Medicine (Baltimore). 2017 May;96(19):e6902. doi: 10.1097/MD.0000000000006902.

Abstract

Shear Wave Elastography (SWE) is an objective and non-invasive method widely used to quantify the tissue solidity. However, there are concerns about the accuracy of the skin SWE results due to the low signal-to-noise ratio (SNR) caused by subcutaneous fat, muscle and bone. This article analyzed the reproducibility of the result for skin SWE and therefore evaluated the availability of SME for skin elasticity involved diseases. Thirty volunteers (mean age: 37 ± 12 years) were selected. SWE were taken on the skin of abdomen and the middle tibia in order to assess the impact of fat, muscle and bone on SWE results. Skin in the area of anterior and lateral tibia marked with seven parallel lines, and each line indicated an identical thickness of the subcutaneous fat from 1-7 mm. Intra-class correlation coefficients (ICC) were used to evaluate the intra-observer and inter-observer reproducibility. The solidity of abdominal skin showed soft and small individual differences (12.4 ± 2.7 kPa), whereas high shear moduli (25-48 kPa) were observed in the skin above tibia and tibialis anterior muscle. When the subcutaneous fat was thicker than 3 mm (≥3), we obtained excellent intra-observer reproducibility (ICC range 0.78-0.98) and inter-observer reproducibility (ICC range 0.75-0.98). The thickness of subcutaneous fat could affect the reproducibility of skin SWE. The further study on skin SWE standardization should be taken.

摘要

剪切波弹性成像(SWE)是一种广泛用于量化组织硬度的客观且非侵入性的方法。然而,由于皮下脂肪、肌肉和骨骼导致的低信噪比(SNR),人们对皮肤SWE结果的准确性存在担忧。本文分析了皮肤SWE结果的可重复性,从而评估了SWE在涉及皮肤弹性疾病中的可用性。选取了30名志愿者(平均年龄:37±12岁)。对腹部皮肤和胫骨中部进行SWE检查,以评估脂肪、肌肉和骨骼对SWE结果的影响。在胫骨前外侧区域的皮肤标记出七条平行线,每条线表示皮下脂肪厚度从1至7毫米,厚度相同。组内相关系数(ICC)用于评估观察者内和观察者间的可重复性。腹部皮肤的硬度显示出柔软且个体差异较小(12.4±2.7千帕),而在胫骨上方和胫骨前肌上方的皮肤中观察到较高的剪切模量(25 - 48千帕)。当皮下脂肪厚度大于3毫米(≥3)时,我们获得了出色的观察者内可重复性(ICC范围0.78 - 0.98)和观察者间可重复性(ICC范围0.75 - 0.98)。皮下脂肪厚度可能会影响皮肤SWE的可重复性。应进一步开展皮肤SWE标准化的研究。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/604f/5428637/12f2d7349283/medi-96-e6902-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验