二维剪切波弹性成像技术的观察者内和观察者间可重复性以及超声经验对获取可靠数据的影响

Intra- and Inter-Observer Reproducibility of a 2-D Shear Wave Elastography Technique and the Impact of Ultrasound Experience in Achieving Reliable Data.

作者信息

Moga Tudor-Voicu, Stepan Ana Maria, Pienar Corina, Bende Felix, Popescu Alina, Șirli Roxana, Dănilă Mirela, Sporea Ioan

机构信息

Department of Gastroenterology and Hepatology, Victor Babeș' University of Medicine and Pharmacy, Timișoara, Romania.

Department of Gastroenterology and Hepatology, Victor Babeș' University of Medicine and Pharmacy, Timișoara, Romania.

出版信息

Ultrasound Med Biol. 2018 Aug;44(8):1627-1637. doi: 10.1016/j.ultrasmedbio.2018.03.029. Epub 2018 May 22.

Abstract

The aim of the study described here was to evaluate the reproducibility of a new shear-wave elastography (SWE) technique, 2-D SWE.GE, and the impact of ultrasound experience in acquiring reliable measurements, as no official recommendations are available for this system. Elastographic measurements (EMs) were obtained in 60 patients using 2-D-SWE with the GE Logiq E9. Three examiners with different levels of experience in ultrasound-based elastography performed 10 valid EMs on each subject: a novice (C.P.) who had no experience in liver elastography and had performed fewer than 50 ultrasound examinations; an elastography expert (A.M.S.) who had more than 1 y of liver elastographic experience in four elastographic methods and had performed more than 1000 ultrasonography examinations; and an ultrasound expert (T.M.) who had no experience in liver elastography and had performed more than 1000 ultrasound examinations. Medians and interquartile ranges were calculated (m/s). We used the inter-class correlation coefficient and Bland-Altman plots with 95% lower and upper limits of agreement to assess the inter- and intra-observer reproducibility of 2-D-SWE.GE measurements. The final study group included 60 patients, 56.7% women and 43.3% men, with a mean age of 33.08 ± 13.83 y and mean body mass index of 22.85 ± 4.04 kg/m. In this group, 73.3% were healthy volunteers and 26.7% had compensated liver cirrhosis. We did not find significant differences between EMs taken by the examiners overall and across study groups. The overall agreement between examiners was excellent: 0.915 (95% confidence interval [CI]: 0.870-0.946). The agreement between the novice and the experienced examiners, respectively, was good to excellent (novice and ultrasound expert: 0.908, 95% CI: 0.846-0.945; novice and elastography expert: 0.885, 95% CI: 0.808-0.931). The intra-observer reproducibility for each of the examiners was excellent; however, the inter-class correlation coefficients were higher for the examiners more experienced in elastography: 0.936 (95% CI: 0.896-0.963) versus 0.966 (95% CI: 0.943-0.980) versus 0.984 (95% CI: 0. 973-0.991). The good ICCs for the median values indicate that 2-D-SWE.GE is a reproducible method. Ultrasound experience did not significantly influence the results.

摘要

本文所述研究的目的是评估一种新的剪切波弹性成像(SWE)技术——二维SWE.GE的可重复性,以及超声经验对获取可靠测量值的影响,因为目前尚无针对该系统的官方建议。使用GE Logiq E9超声仪的二维SWE对60例患者进行了弹性成像测量(EMs)。三名在基于超声的弹性成像方面经验水平不同的检查者对每个受试者进行了10次有效的EMs测量:一名新手(C.P.),没有肝脏弹性成像经验,超声检查次数少于50次;一名弹性成像专家(A.M.S.),在四种弹性成像方法方面有超过1年的肝脏弹性成像经验,超声检查次数超过1000次;一名超声专家(T.M.),没有肝脏弹性成像经验,超声检查次数超过1000次。计算中位数和四分位间距(m/s)。我们使用组内相关系数和带有95%一致性下限和上限的Bland-Altman图来评估二维SWE.GE测量的观察者间和观察者内的可重复性。最终研究组包括60例患者,女性占56.7%,男性占43.3%,平均年龄为33.08±13.83岁,平均体重指数为22.85±4.04kg/m²。在该组中,73.3%为健康志愿者,26.7%患有代偿期肝硬化。我们没有发现检查者总体以及在各个研究组之间所进行的EMs测量有显著差异。检查者之间的总体一致性非常好:0.915(95%置信区间[CI]:0.870 - 0.946)。新手与有经验的检查者之间的一致性分别为良好到非常好(新手与超声专家:0.908,95% CI:0.846 - 0.945;新手与弹性成像专家:0.885,95% CI:0.808 - 0.931)。每位检查者的观察者内可重复性都非常好;然而,在弹性成像方面经验更丰富的检查者的组内相关系数更高:0.936(95% CI:0.896 - 0.963)对比0.966(95% CI:0.943 - 0.980)对比0.984(95% CI:0.973 - 0.991)。中位数的良好组内相关系数表明二维SWE.GE是一种可重复的方法。超声经验对结果没有显著影响。