二维剪切波弹性成像评估肝脏的可重复性:与健康志愿者的点剪切波弹性成像的直接比较

Reproducibility of 2-Dimensional Shear Wave Elastography Assessment of the Liver: A Direct Comparison With Point Shear Wave Elastography in Healthy Volunteers.

作者信息

Fang Cheng, Konstantatou Eleni, Romanos Odyssefs, Yusuf Gibran T, Quinlan Daniel J, Sidhu Paul S

机构信息

Department of Radiology, King's College Hospital, London, United Kingdom.

出版信息

J Ultrasound Med. 2017 Aug;36(8):1563-1569. doi: 10.7863/ultra.16.07018. Epub 2017 Apr 3.

Abstract

OBJECTIVES

Two-dimensional shear wave elastography (2D-SWE) imaging for the noninvasive assessment of tissue stiffness was assessed for reproducibility in healthy volunteers in quantifying liver elasticity, compared with an established point shear wave elastography (p-SWE) technique also known as virtual touch quantification (VTQ) (SIEMENS).

METHODS

Eleven healthy volunteers were examined by four experienced operators on two occasions, separated by two weeks (sessions A and B). Ten 2D-SWE using LOGIQ E9 and p-SWE measurements using VTQ (in meters per second) were consecutively taken from deep portions of liver segments 5 or 6 away from vascular structures, using standard techniques. Inter- and intra-observer agreement was assessed by intraclass coefficient (ICC).

RESULTS

A total of 880 2D-SWE and p-SWE velocities were recorded. Mean values from the four operators ranged between 1.188 and 1.196 m/s for 2D-SWE and 1.170 to 1.207 m/s for p-SWE. Interobserver reproducibility was good for both sessions with ICCs of 0.88 and 0.93 (2D-SWE) and 0.87 and 0.93 (p-SWE). The overall intra-operator reproducibility between sessions A and B was good for both p-SWE and 2D-SWE with ICC of 0.87 and 0.83, respectively. For inter- and intra-observer variability, the ICC was more than or equal to 0.71, indicating that the results were reliable. There was a strong and significant correlation between the 2D-SWE and p-SWE measurements (r = 0.87, P = .0006), but their velocities did not agree equally across different velocities.

CONCLUSIONS

Two-dimensional SWE using LOGIQ E9 is a reliable and reproducible method for measuring elasticity in healthy volunteers and has a similar degree of reliability as p-SWE using VTQ, but absolute measurements from the two techniques should not be used interchangeably.

摘要

目的

评估二维剪切波弹性成像(2D-SWE)用于无创评估组织硬度在健康志愿者中量化肝脏弹性的可重复性,与一种成熟的点剪切波弹性成像(p-SWE)技术(也称为虚拟触诊定量(VTQ),西门子公司)进行比较。

方法

11名健康志愿者由4名经验丰富的操作人员分两次进行检查,两次检查间隔两周(A组和B组)。使用标准技术,从肝脏第5或6段远离血管结构的深部连续采集10次使用LOGIQ E9的2D-SWE和使用VTQ(以米每秒为单位)的p-SWE测量值。通过组内相关系数(ICC)评估观察者间和观察者内的一致性。

结果

共记录了880个2D-SWE和p-SWE速度值。四位操作人员的2D-SWE平均值在1.188至1.196米/秒之间,p-SWE平均值在1.170至1.207米/秒之间。观察者间的可重复性在两组中均良好,2D-SWE的ICC分别为0.88和0.93,p-SWE的ICC分别为0.87和0.93。A组和B组之间操作人员总体的可重复性对于p-SWE和2D-SWE均良好,ICC分别为0.87和0.83。对于观察者间和观察者内的变异性,ICC大于或等于0.71,表明结果可靠。2D-SWE和p-SWE测量值之间存在强且显著的相关性(r = 0.87,P = 0.0006),但它们在不同速度下的速度并不完全一致。

结论

使用LOGIQ E9的二维SWE是一种用于测量健康志愿者弹性的可靠且可重复的方法,与使用VTQ的p-SWE具有相似程度的可靠性,但这两种技术的绝对测量值不应互换使用。