Magn Reson Med Sci. 2018 Jan 10;17(1):35-41. doi: 10.2463/mrms.mp.2016-0092. Epub 2017 May 18.
To evaluate the effect of a composite radiofrequency (RF) pulse on saturation recovery (SR) myocardial T mapping using a 3T MR system.
Phantom and in vivo studies were performed with a clinical 3T MR scanner. Accuracy and reproducibility of the SR T mapping using conventional and composite RF pulses were first compared in phantom experiments. An in vivo study was performed of 10 healthy volunteers who were imaged with conventional and composite RF pulse methods twice each. In vivo reproducibility of myocardial T value and the inter-segment variability were assessed.
The phantom study revealed significant differences in the mean T values between the two methods, and the reproducibility for the composite RF pulse was significantly smaller than that for the conventional RF pulse. For both methods, the correlations of the reference and measured T values were excellent (r = 0.97 and 0.98 for conventional and composite RF pulses, respectively). The in vivo study showed that the mean T value for composite RF pulse was slightly lower than that for conventional RF pulse, but this difference was not significant (P = 0.06). The inter-segment variability for the composite RF pulse was significantly smaller than that for conventional RF pulse (P < 0.01). Inter-scan correlations of T measurements of the first and second scans were highly and weakly correlated to composite RF pulses (r = 0.83 and 0.29, respectively).
SR T mapping using composite RF pulse provides accurate quantification of T values and can lessen measurement variability and enable reproducible T measurements.
评估使用 3T MR 系统的复合射频(RF)脉冲对饱和恢复(SR)心肌 T 映射的影响。
使用临床 3T MR 扫描仪进行了体模和体内研究。首先在体模实验中比较了传统和复合 RF 脉冲对 SR T 映射的准确性和可重复性。对 10 名健康志愿者进行了两次常规和复合 RF 脉冲方法的体内研究。评估了心肌 T 值的体内重复性和节段间可变性。
体模研究显示两种方法之间的平均 T 值存在显著差异,复合 RF 脉冲的可重复性明显小于常规 RF 脉冲。对于两种方法,参考和测量 T 值之间的相关性均非常好(常规 RF 脉冲和复合 RF 脉冲的 r 值分别为 0.97 和 0.98)。体内研究表明,复合 RF 脉冲的平均 T 值略低于常规 RF 脉冲,但差异无统计学意义(P = 0.06)。复合 RF 脉冲的节段间可变性明显小于常规 RF 脉冲(P < 0.01)。首次和第二次扫描的 T 值测量的扫描间相关性与复合 RF 脉冲具有高度和弱相关性(r 值分别为 0.83 和 0.29)。
使用复合 RF 脉冲的 SR T 映射可准确量化 T 值,并减少测量变异性,实现可重复的 T 值测量。