Poster - Thur Eve - 10: Long term stability of VMAT quality assurance parameters using an EPID.

作者信息

Pekar J, Diamond K R

机构信息

Department of Medical Physics, Juravinski Cancer Centre, Hamilton, ON, Canada.

McMaster University, Hamilton, ON, Canada.

出版信息

Med Phys. 2012 Jul;39(7Part2):4626. doi: 10.1118/1.4740118.

Abstract

The rapidly growing use of volumetric modulated arc therapy (VMAT) treatments in radiation therapy calls for a quantitative, automated, and reliable quality assurance (QA) procedure that can be used routinely in the clinical setting. In this work, we present a series VMAT QA procedures used to assess dynamic multi-leaf collimator (MLC) positional accuracy, variable dose-rate accuracy, and MLC leaf speed accuracy. The QA procedures were performed using amorphous silicon electronic portal imaging devices (EPID) to determine the long term stability of the measured parameters on two Varian linear accelerators. The measurements were repeated weekly on both linear accelerators for a period of three months and the EPID images were analyzed using custom Matlab software. The results of the picket fence tests indicate that MLC leaf positions can be identified to within 0.11 mm and 0.15 mm for static gantry delivery and VMAT delivery respectively. In addition, the dose-rate, gantry speed and MLC leaf speed tests both show very good stability over the measurement period. The measurements thus far, suggest that a number of the dosimetry tests may be suitable for quarterly QA for Varian iX and Trilogy linacs. However, additional measurements are required to confirm the frequency with which each test is required for safe and reliable VMAT delivery at our centre.

摘要

容积调强弧形治疗(VMAT)在放射治疗中的应用迅速增加,这就需要一种可在临床环境中常规使用的定量、自动化且可靠的质量保证(QA)程序。在这项工作中,我们展示了一系列用于评估动态多叶准直器(MLC)位置精度、可变剂量率精度和MLC叶片速度精度的VMAT QA程序。QA程序使用非晶硅电子射野影像装置(EPID)来确定在两台瓦里安直线加速器上测量参数的长期稳定性。在两台直线加速器上每周重复进行测量,为期三个月,并使用定制的Matlab软件分析EPID图像。栅栏测试结果表明,对于静态机架照射和VMAT照射,MLC叶片位置分别可精确到0.11毫米和0.15毫米以内。此外,剂量率、机架速度和MLC叶片速度测试在测量期间均显示出非常好的稳定性。迄今为止的测量结果表明,一些剂量测定测试可能适用于瓦里安iX和Trilogy直线加速器的季度QA。然而,还需要进行额外的测量,以确定在我们中心安全可靠地进行VMAT照射时每项测试所需的频率。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验