SU-E-T-112: Experimental Characterization of a Novel Thermal Reservoir for Consistent and Accurate Annealing of High-Sensitivity TLDs.

作者信息

Donahue W, Bongiorni P, Hearn R, Rodgers J, Nath R, Chen Z

机构信息

Yale University School of Medicine, New Haven, CT.

Theragenics Corporation, Atlanta, GA.

出版信息

Med Phys. 2012 Jun;39(6Part11):3728-3729. doi: 10.1118/1.4735170.

Abstract

PURPOSE

To develop and characterize a novel thermal reservoir for consistent and accurate annealing of high-sensitivity thermoluminescence dosimeters (TLD-100H) for dosimetry of brachytherapy sources.

METHODS

The sensitivity of TLD-100H is about 18 times that of TLD-100 which has clear advantages in for interstitial brachytherapy sources. However, the TLD-100H requires a short high temperature annealing cycle (15 min.) and opening and closing the oven door causes significant temperature fluctuations leading to unreliable measurements. A new thermal reservoir made of aluminum alloy was developed to provide stable temperature environment in a standard hot air oven. The thermal reservoir consisted of a 20 cm × 20 cm × 8 cm Al block with a machine-milled chamber in the middle to house the aluminum TLD holding tray. The thermal reservoir was placed inside the oven until it reaches thermal equilibrium with oven chamber. The temperatures of the oven chamber, heat reservoir, and TLD holding tray were monitored by two independent thermo-couples which interfaced digitally to a control computer. A LabView interface was written for monitoring and recording the temperatures in TLD holding tray, the thermal reservoir, and oven chamber. The temperature profiles were measured as a function of oven-door open duration. The settings for oven chamber temperature and oven door open-close duration were optimized to achieve a stable temperature of 240 0C in the TLD holding tray. Complete temperature profiles of the TLD annealing tray over the entire annealing process were obtained.

RESULTS

A LabView interface was written for monitoring and recording the temperatures in TLD holding The use of the thermal reservoir has significantly reduced the temperature fluctuations caused by the opening of oven door when inserting the TLD holding tray into the oven chamber. It has enabled consistent annealing of high-sensitivity TLDs.

CONCLUSIONS

A comprehensive characterization of a custom-built novel thermal reservoir for annealing high-sensitivity TLD has been carried out. It enabled consistent and accurate annealing of high- sensitivity TLDs which could significantly improve the efficiency of brachytherapy source characterizations. Supported in part by NIH grant R01-CA134627.

摘要

目的

开发并表征一种新型热库,用于对高灵敏度热释光剂量计(TLD-100H)进行一致且准确的退火处理,以用于近距离放射治疗源的剂量测定。

方法

TLD-100H的灵敏度约为TLD-100的18倍,这在间质近距离放射治疗源方面具有明显优势。然而,TLD-100H需要短时间的高温退火周期(15分钟),并且打开和关闭烤箱门会导致显著的温度波动,从而导致测量结果不可靠。开发了一种由铝合金制成的新型热库,以在标准热风烤箱中提供稳定的温度环境。热库由一个20厘米×20厘米×8厘米的铝块组成,中间有一个机械加工的腔室,用于放置铝制TLD托盘。将热库放置在烤箱内,直到它与烤箱腔室达到热平衡。烤箱腔室、热库和TLD托盘的温度由两个独立的热电偶监测,这两个热电偶通过数字方式与一台控制计算机连接。编写了一个LabView接口,用于监测和记录TLD托盘、热库和烤箱腔室中的温度。测量了温度曲线与烤箱门打开持续时间的函数关系。优化了烤箱腔室温度和烤箱门打开-关闭持续时间的设置,以在TLD托盘中实现240℃的稳定温度。获得了TLD退火托盘在整个退火过程中的完整温度曲线。

结果

编写了一个LabView接口,用于监测和记录TLD中的温度。使用热库显著减少了将TLD托盘插入烤箱腔室时因打开烤箱门而引起的温度波动。它能够对高灵敏度TLD进行一致的退火处理。

结论

对用于高灵敏度TLD退火的定制新型热库进行了全面表征。它能够对高灵敏度TLD进行一致且准确的退火处理,这可以显著提高近距离放射治疗源表征的效率。部分得到美国国立卫生研究院R01-CA134627资助。