Suppr超能文献

低剂量场定量热释光剂量测定中的仪器和剂量计尺寸伪影

Instrumentation and dosimeter-size artifacts in quantitative thermoluminescence dosimetry of low-dose fields.

作者信息

Meigooni A S, Mishra V, Panth H, Williamson J

机构信息

Washington University School of Medicine, Mallinckrodt Institute of Radiology, St. Louis, Missouri 63110, USA.

出版信息

Med Phys. 1995 May;22(5):555-61. doi: 10.1118/1.597555.

Abstract

Thermoluminescence dosimetry is extensively used for quantitative dose measurements in various irradiation fields such as dosimetry of brachytherapy sources. In this application, small doses on the order of 0.5 cGy must be accurately measured, which requires careful control of instrumentation, energy-dependence, and nonlinearity of detector response. Several investigators have observed the presence of some undesirable signals when the thermoluminescent dosimeters (TLDs) were read without any nitrogen gas flow in the TLD reader. Others have indicated that the "prereadout" annealing technique is the same as the "preirradiation" technique for doses above 10 cGy, but they have not extended their study to lower doses. The goal of this study is to investigate dependence of sensitivity and linearity of the TLD response to the flow of nitrogen gas in the TLD reader at low dose level, annealing technique, and TLD size. The effect of nitrogen flow sensitivity and linearity of two different sizes of lithium fluoride TLD-100 chips has been studied. Our data indicate a large standard deviation of TLD sensitivity, up to a factor of 2, when TLDs were read without nitrogen gas flow in the TLD reader. In addition, a large deviation from linearity was observed for doses below 5 cGy. When the reading-chamber was purged with nitrogen gas, dispersion of the responses of the TLDs that were exposed to the same dose fell to within 5%. At precision levels of 2% and 5%, the low dose limits are 1 cGy and 0.5 cGy, respectively, for large chips and 15 cGy and 1 cGy for small chips, if TLDs are read with nitrogen gas flow in the TLD reader.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

热释光剂量测定法广泛应用于各种辐射场的定量剂量测量,如近距离治疗源的剂量测定。在这种应用中,必须精确测量约0.5 cGy量级的小剂量,这需要仔细控制仪器、能量依赖性以及探测器响应的非线性。几位研究者观察到,当热释光剂量计(TLD)在TLD读数器中无氮气流动的情况下读取时,会出现一些不良信号。其他人指出,对于高于10 cGy的剂量,“预读出”退火技术与“预辐照”技术相同,但他们尚未将研究扩展到更低剂量。本研究的目的是在低剂量水平、退火技术和TLD尺寸方面,研究TLD响应的灵敏度和线性对TLD读数器中氮气流动的依赖性。已经研究了两种不同尺寸的氟化锂TLD - 100芯片的氮气流动对灵敏度和线性的影响。我们的数据表明,当TLD在TLD读数器中无氮气流动的情况下读取时,TLD灵敏度的标准偏差很大,高达2倍。此外,对于低于5 cGy的剂量,观察到与线性有很大偏差。当用氮气吹扫读数室时,暴露于相同剂量的TLD响应的分散度降至5%以内。如果TLD在TLD读数器中有氮气流动的情况下读取,对于大芯片,在精度水平为2%和5%时,低剂量限值分别为1 cGy和0.5 cGy,对于小芯片则分别为15 cGy和1 cGy。(摘要截断于250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验