Silkwood J, Matthews K, Shikhaliev P
Louisiana State University, Baton Rouge, LA.
Med Phys. 2012 Jun;39(6Part24):3914. doi: 10.1118/1.4735976.
Photon counting spectral breast CT is feasible in part due to using an adaptive filter. An adaptive filter provides flat x-ray intensity profile and constant x-ray energy spectrum across detector surface, decreases required detector count rate, and eliminates beam hardening artifacts. However, the altered x-ray exposure profiles at the breast and detector surface may influence the distribution of CT noise, CT numbers, and contrast to noise ratio (CNR) across the CT images. The purpose of this work was to investigate these effects.
Images of a CT phantom with and without adaptive filter were simulated at 60kVp, 90kVp, and 120kVp tube voltages and 660 mR total skin exposure. The CT phantom with water content had 14cm diameter, contrast elements representing adipose tissue and 2.5mg/cc iodine contrast located at 1cm, 3.5cm, and 6cm from center of the phantom. The CT numbers, CT noise, and CNR were measured at multiple locations for several filter/exposure combinations: (1)without adaptive filter for 660mR skin exposure; (2)with adaptive filter for 660mR skin exposure along central axis (mean skin exposure across the breast was <660mR); and (3)with adaptive filter for scaled exposure (mean skin exposure was 660mR).
Beam hardening (cupping) artifacts had 47HU magnitude without adaptive filter but were eliminated with adaptive filter. CNR of contrast elements was comparable for (1) and (2) over central parts but was higher by 20-30% for (1) near the edge of the phantom. CNR was higher by 20-30% in (3) as compared to (2) over central parts and comparable near the edges.
The adaptive filter provided: uniform distribution of CT noise, CNR, and CT numbers across CT images; comparable or better CNR with no dose penalty to the breast; and eliminated beam hardening artifacts.
光子计数光谱乳腺CT可行,部分原因是采用了自适应滤波器。自适应滤波器可在探测器表面提供平坦的X射线强度分布和恒定的X射线能谱,降低所需的探测器计数率,并消除束硬化伪影。然而,乳腺和探测器表面X射线曝光分布的改变可能会影响CT图像上CT噪声、CT值以及对比度噪声比(CNR)的分布。本研究旨在探究这些影响。
在60kVp、90kVp和120kVp管电压以及660mR总皮肤曝光量条件下,模拟了有无自适应滤波器时CT体模的图像。含水的CT体模直径为14cm,有代表脂肪组织的对比元素以及位于距体模中心1cm、3.5cm和6cm处的2.5mg/cc碘对比剂。针对几种滤波器/曝光组合,在多个位置测量CT值、CT噪声和CNR:(1)无自适应滤波器,皮肤曝光量为660mR;(2)有自适应滤波器,沿中心轴皮肤曝光量为660mR(乳腺平均皮肤曝光量<660mR);(3)有自适应滤波器,按比例曝光(平均皮肤曝光量为660mR)。
无自适应滤波器时束硬化(杯状)伪影大小为47HU,但使用自适应滤波器后被消除。对比元素的CNR在(1)和(2)的中心部分相当,但在(1)中靠近体模边缘处比(2)高20 - 30%。(3)中CNR在中心部分比(2)高20 - 30%,在边缘处相当。
自适应滤波器实现了:CT噪声、CNR和CT值在CT图像上的均匀分布;在不增加乳腺剂量的情况下具有相当或更好的CNR;并消除了束硬化伪影。