Li Na, Hu Bo, Mi Fanglin, Song Jinlin
Department of Stomatology, Affiliated Hospital of North Sichuan Medical College, Nanchong, Sichuan 637000, P.R. China.
Department of Orthodontics, Stomatological Hospital Affiliated to Chongqing Medical University, Chongqing 401147, P.R. China.
Exp Ther Med. 2017 May;13(5):2451-2455. doi: 10.3892/etm.2017.4278. Epub 2017 Mar 29.
The present study was conducted to evaluate the three-dimensional (3D) cephalometry accuracy of cone-beam computed tomography (CBCT). A total of 40 newly diagnosed orthodontic patients (including 18 males and 22 females; age range, 12-18 years) were subjected to CBCT scanning and X-ray imaging in order to obtain lateral cephalograms. The 3D CBCT cephalograms were reconstructed and measured with 5.1 software, and compared with the results from the conventional 2D lateral cephalograms. Measurements in the two images were performed twice using the Steiner and Tweed standards and a single analyzer paired t-test was used to analyze the differences between the two cephalometric methods. The results indicated that the two methods showed significant differences in all 12 angle and 5 linear measurements (P<0.05). These findings indicated that CBCT is a more accurate technique compared with the conventional 2D method. In conclusion, CBCT may provide diagnostic and treatment information for maxillofacial deformities by using fast computer-aided analysis platform.
本研究旨在评估锥形束计算机断层扫描(CBCT)的三维(3D)头影测量准确性。总共40例新诊断的正畸患者(包括18例男性和22例女性;年龄范围为12至18岁)接受了CBCT扫描和X线成像,以获得头颅侧位片。使用5.1软件重建并测量3D CBCT头颅侧位片,并与传统二维头颅侧位片的结果进行比较。使用Steiner和Tweed标准在两张图像上进行两次测量,并使用单分析器配对t检验分析两种头影测量方法之间的差异。结果表明,两种方法在所有12个角度和5个线性测量中均显示出显著差异(P<0.05)。这些发现表明,与传统二维方法相比,CBCT是一种更准确的技术。总之,CBCT可通过使用快速计算机辅助分析平台为颌面部畸形提供诊断和治疗信息。