Suppr超能文献

正畸头影测量在超低剂量-低剂量 CBCT 重建头影测量片中的精度。

Precision of orthodontic cephalometric measurements on ultra low dose-low dose CBCT reconstructed cephalograms.

机构信息

Department of Orthodontics, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands.

Department of Rehabilitation and Department of Oral and Maxillofacial Surgery, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands.

出版信息

Clin Oral Investig. 2022 Feb;26(2):1543-1550. doi: 10.1007/s00784-021-04127-9. Epub 2021 Aug 28.

Abstract

OBJECTIVES

To analyze differences in variation of orthodontic diagnostic measurements on lateral cephalograms reconstructed from ultra low dose-low dose (ULD-LD) cone beam computed tomography (CBCT) scans (RLC) as compared to variation of measurements on standard lateral cephalograms (SLC), and to determine if it is justifiable to replace a traditional orthodontic image set for an ULD-LD CBCT with a reconstructed lateral cephalogram.

MATERIAL AND METHODS

ULD-LD CBCT images and SLCs were made of forty-three dry human skulls. From the ULD-LD CBCT dataset, a lateral cephalogram was reconstructed (RLC). Cephalometric landmarks (13 skeletal and 7 dental) were identified on both SLC and RLC twice in two sessions by two calibrated observers. Thirteen cephalometric variables were calculated. Variations of measurements, expressed as standard deviations of the 4 measurements on SLC and RLC, were analyzed using a paired sample t-test. Differences in the number of observations deviating ≥ 2.0 mm or degrees from the grand mean between SLC and RLC were analyzed using a McNemar test.

RESULTS

Mean SDs for 7 out of 13 variables were significantly smaller for SLCs than those for RLCs, but differences were small. For 9 out of 13 variables, there was no significant difference between SLC and RLC for the number of measurements outside the range of 2 mm or degrees.

CONCLUSIONS

Based on the lower radiation dose and the small differences in variation in cephalometric measurements on reconstructed LC compared to standard dose LC, ULD-LD CBCT with reconstructed LC should be considered for orthodontic diagnostic purposes.

CLINICAL RELEVANCE

ULD-LD CBCT with reconstructed LC should be considered for orthodontic purposes.

摘要

目的

分析超低剂量-低剂量(ULD-LD)锥形束 CT(CBCT)扫描重建的侧位头颅侧位片(RLC)与标准侧位头颅侧位片(SLC)上正畸诊断测量值的变化差异,并确定是否可以用重建的侧位头颅侧位片代替传统的 ULD-LD CBCT 正畸图像集。

材料与方法

对 43 个干燥人颅骨进行 ULD-LD CBCT 和 SLC 成像。从 ULD-LD CBCT 数据集重建了一个侧位头颅片(RLC)。由两位经过校准的观察者在两个会话中两次在 SLC 和 RLC 上识别 13 个骨骼和 7 个牙齿的头影测量标志点。计算了 13 个头影测量变量。使用配对样本 t 检验分析表示 SLC 和 RLC 上 4 次测量的标准差的测量值变化。使用 McNemar 检验分析 SLC 和 RLC 之间偏离总平均值≥2.0 毫米或度数的观测值数量的差异。

结果

13 个变量中的 7 个变量的平均标准差 SLC 明显小于 RLC,但差异较小。对于 13 个变量中的 9 个变量,SLC 和 RLC 之间在 2 毫米或度数范围内的测量值没有显著差异。

结论

基于较低的辐射剂量和重建 LC 上与标准剂量 LC 相比头影测量值变化的微小差异,应考虑将 ULD-LD CBCT 与重建 LC 用于正畸诊断目的。

临床相关性

应考虑将 ULD-LD CBCT 与重建 LC 用于正畸目的。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/39ff/8816531/682aae763867/784_2021_4127_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验