Clin Oral Investig. 2022 Feb;26(2):1543-1550. doi: 10.1007/s00784-021-04127-9. Epub 2021 Aug 28.
To analyze differences in variation of orthodontic diagnostic measurements on lateral cephalograms reconstructed from ultra low dose-low dose (ULD-LD) cone beam computed tomography (CBCT) scans (RLC) as compared to variation of measurements on standard lateral cephalograms (SLC), and to determine if it is justifiable to replace a traditional orthodontic image set for an ULD-LD CBCT with a reconstructed lateral cephalogram.
ULD-LD CBCT images and SLCs were made of forty-three dry human skulls. From the ULD-LD CBCT dataset, a lateral cephalogram was reconstructed (RLC). Cephalometric landmarks (13 skeletal and 7 dental) were identified on both SLC and RLC twice in two sessions by two calibrated observers. Thirteen cephalometric variables were calculated. Variations of measurements, expressed as standard deviations of the 4 measurements on SLC and RLC, were analyzed using a paired sample t-test. Differences in the number of observations deviating ≥ 2.0 mm or degrees from the grand mean between SLC and RLC were analyzed using a McNemar test.
Mean SDs for 7 out of 13 variables were significantly smaller for SLCs than those for RLCs, but differences were small. For 9 out of 13 variables, there was no significant difference between SLC and RLC for the number of measurements outside the range of 2 mm or degrees.
Based on the lower radiation dose and the small differences in variation in cephalometric measurements on reconstructed LC compared to standard dose LC, ULD-LD CBCT with reconstructed LC should be considered for orthodontic diagnostic purposes.
ULD-LD CBCT with reconstructed LC should be considered for orthodontic purposes.
分析超低剂量-低剂量(ULD-LD)锥形束 CT(CBCT)扫描重建的侧位头颅侧位片(RLC)与标准侧位头颅侧位片(SLC)上正畸诊断测量值的变化差异,并确定是否可以用重建的侧位头颅侧位片代替传统的 ULD-LD CBCT 正畸图像集。
对 43 个干燥人颅骨进行 ULD-LD CBCT 和 SLC 成像。从 ULD-LD CBCT 数据集重建了一个侧位头颅片(RLC)。由两位经过校准的观察者在两个会话中两次在 SLC 和 RLC 上识别 13 个骨骼和 7 个牙齿的头影测量标志点。计算了 13 个头影测量变量。使用配对样本 t 检验分析表示 SLC 和 RLC 上 4 次测量的标准差的测量值变化。使用 McNemar 检验分析 SLC 和 RLC 之间偏离总平均值≥2.0 毫米或度数的观测值数量的差异。
13 个变量中的 7 个变量的平均标准差 SLC 明显小于 RLC,但差异较小。对于 13 个变量中的 9 个变量,SLC 和 RLC 之间在 2 毫米或度数范围内的测量值没有显著差异。
基于较低的辐射剂量和重建 LC 上与标准剂量 LC 相比头影测量值变化的微小差异,应考虑将 ULD-LD CBCT 与重建 LC 用于正畸诊断目的。
应考虑将 ULD-LD CBCT 与重建 LC 用于正畸目的。