超声向量血流成像——会成为评估用于血液透析的动静脉内瘘的一种新工具吗?

Ultrasound Vector Flow Imaging - could be a new tool in evaluation of arteriovenous fistulas for hemodialysis?

作者信息

Fiorina Ilaria, Raciti Maria Vittoria, Goddi Alfredo, Cantisani Vito, Bortolotto Chandra, Chu Shane, Calliada Fabrizio

机构信息

Radiology Department, Fondazione IRCCS Policlinico San Matteo, Università degli Studi di Pavia, Pavia - Italy.

Centro Medico SME-Diagnostica per Immagini, Varese - Italy.

出版信息

J Vasc Access. 2017 Jul 14;18(4):284-289. doi: 10.5301/jva.5000721. Epub 2017 May 24.

Abstract

INTRODUCTION

We report the use of a new ultrasound technique to evaluate the axial and lateral components of a complex flow in the arteriovenous fistula (AVF). Vector Flow Imaging (VFI) allows to identify different components of the flow in every direction, even orthogonal to the flow streamline, represented by many single vectors. VFI could help to identify flow alterations in AVF, probably responsible for its malfunction.

METHODS

From February to June 2016, 14 consecutive patients with upper-limb AVF were examined with a Resona 7 (Mindray, Shenzhen, China) ultrasound scanner equipped with VFI. An analysis of mean velocity, angular direction and mean number of vectors impacting the vessel wall was carried out. We also identified main flow patterns present in the arterial side, into the venous aneurysm and in correspondence of significant stenosis.

RESULTS

A disturbed flow with the presence of vectors directed against the vessel walls was found in 9/14 patients (64.28%): in correspondence of the iuxta-anastomotic venous side (4/9; 44.4%), into the venous aneurysmal tracts (3/9; 33.3%) and in concomitance of stenosis (2/9; 22.2%). The mean velocity of the vectors was around 20-25 cm/s, except in presence of stenosis, where the velocities were much higher (45-50 cm/s). The vectors directed against the vessel walls presented high angle attack (from 45° to 90°, with a median angular deviation 65°).

CONCLUSIONS

VFI was confirmed to be an innovative and intuitive imaging technology to study the flow complexity in the arteriovenous fistulas.

摘要

引言

我们报告了一种新的超声技术用于评估动静脉内瘘(AVF)中复杂血流的轴向和侧向成分。矢量血流成像(VFI)能够识别各个方向上血流的不同成分,甚至是与血流流线正交的方向,这些方向由许多单个矢量表示。VFI有助于识别AVF中的血流改变,这可能是其功能障碍的原因。

方法

2016年2月至6月,对14例连续的上肢AVF患者使用配备VFI的Resona 7(深圳迈瑞)超声扫描仪进行检查。对平均速度、角度方向以及撞击血管壁的矢量平均数量进行了分析。我们还确定了动脉侧、静脉瘤内以及明显狭窄处存在的主要血流模式。

结果

在9/14例患者(64.28%)中发现了血流紊乱,存在指向血管壁的矢量:在吻合口旁静脉侧(4/9;44.4%)、静脉瘤样管道内(3/9;33.3%)以及合并狭窄处(2/9;22.2%)。矢量的平均速度约为20 - 25厘米/秒,但在存在狭窄的情况下,速度要高得多(45 - 50厘米/秒)。指向血管壁的矢量呈现高角度攻击(45°至90°,中位角度偏差65°)。

结论

VFI被证实是一种用于研究动静脉内瘘血流复杂性的创新且直观的成像技术。