Lind Emelie, Knutsson Linda, Kämpe Robin, Ståhlberg Freddy, Wirestam Ronnie
Department of Medical Radiation Physics, Lund University, Barngatan 4, 22185, Lund, Sweden.
Department of Radiology (Adjunct), Johns Hopkins School of Medicine, Baltimore, USA.
MAGMA. 2017 Dec;30(6):555-566. doi: 10.1007/s10334-017-0637-9. Epub 2017 Jun 19.
One major issue in dynamic susceptibility contrast MRI (DSC-MRI) is to accurately determine contrast agent (CA) concentration, since T2* relaxivity in vivo is generally unknown and varies between blood and tissue. In this study, quantitative susceptibility mapping (QSM) was used for quantification of CA concentration.
A DSC-MRI protocol, including phase data acquisition, was applied to 20 healthy volunteers in a test-retest study. By selecting a CSF reference region of interest (ROI), the values of all QSM images were shifted to show no CA-induced change in CSF. CA concentration and cerebral blood volume (CBV) were estimated using shifted QSM data. CSF reference ROI optimization was evaluated by investigation of CBV repeatability. The CBV age dependence was analysed and tissue T2* relaxivity was estimated.
The best repeatability of CBV, using an optimal CSF reference ROI, showed test-versus-retest correlations of r = 0.81 and r = 0.91 for white and grey matter, respectively. A slight CBV decrease with age was observed, and the estimated in vivo T2* relaxivity was 85 mMs.
Provided that a carefully selected CSF reference ROI is used to shift QSM image values, susceptibility information can be used to estimate concentration of contrast agent and to calculate CBV.
在动态磁敏感对比增强磁共振成像(DSC-MRI)中,一个主要问题是准确测定造影剂(CA)浓度,因为体内的T2*弛豫率通常未知且在血液和组织之间存在差异。在本研究中,定量磁敏感成像(QSM)被用于CA浓度的定量分析。
在一项重测研究中,将包括相位数据采集的DSC-MRI方案应用于20名健康志愿者。通过选择一个脑脊液(CSF)参考感兴趣区(ROI),所有QSM图像的值被偏移,以显示CSF中无CA诱导的变化。使用偏移后的QSM数据估计CA浓度和脑血容量(CBV)。通过研究CBV的重复性来评估CSF参考ROI的优化。分析了CBV与年龄的相关性,并估计了组织T2*弛豫率。
使用最佳CSF参考ROI时,CBV的最佳重复性显示,白质和灰质的重测相关性分别为r = 0.81和r = 0.91。观察到CBV随年龄略有下降,估计的体内T2*弛豫率为85 mM/s。
如果使用精心选择的CSF参考ROI来偏移QSM图像值,则磁敏感信息可用于估计造影剂浓度并计算CBV。