Photon detection efficiency of laboratory-based x-ray phase contrast imaging techniques for mammography: a Monte Carlo study.

作者信息

Saghamanesh S, Aghamiri S M, Kamali-Asl A, Yashiro W

机构信息

Department of Medical Radiation Engineering, Shahid Beheshti University, Tehran 1983969411, Iran.

出版信息

Phys Med Biol. 2017 Sep 1;62(18):7394-7406. doi: 10.1088/1361-6560/aa7a92.

Abstract

An important challenge in real-world biomedical applications of x-ray phase contrast imaging (XPCI) techniques is the efficient use of the photon flux generated by an incoherent and polychromatic x-ray source. This efficiency can directly influence dose and exposure time and ideally should not affect the superior contrast and sensitivity of XPCI. In this paper, we present a quantitative evaluation of the photon detection efficiency of two laboratory-based XPCI methods, grating interferometry (GI) and coded-aperture (CA). We adopt a Monte Carlo approach to simulate existing prototypes of those systems, tailored for mammography applications. Our simulations were validated by means of a simple experiment performed on a CA XPCI system. Our results show that the fraction of detected photons in the standard energy range of mammography are about 1.4% and 10% for the GI and CA techniques, respectively. The simulations indicate that the design of the optical components plays an important role in the higher efficiency of CA compared to the GI method. It is shown that the use of lower absorbing materials as the substrates for GI gratings can improve its flux efficiency by up to four times. Along similar lines, we also show that an optimized and compact configuration of GI could lead to a 3.5 times higher fraction of detected counts compared to a standard and non-optimised GI implementation.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验