Shin Mirim, Patton Raymond, Mahar Trevor, Ireland Angus, Swan Paul, Chow Chin Moi
Exercise, Heath & Performance Research Group, Faculty of Health Sciences, The University of Sydney, Australia.
Exercise, Heath & Performance Research Group, Faculty of Health Sciences, The University of Sydney, Australia.
Physiol Behav. 2017 Oct 1;179:208-212. doi: 10.1016/j.physbeh.2017.06.019. Epub 2017 Jun 27.
Accurate relative humidity (RH) measurement is demanded in studies of thermal comfort. Thermal discomfort occurs when the near-to-skin temperature or RH is outside of the thermal comfort zone. The Hygrochron, a small wireless device which measures both temperature and RH, would be suitable and convenient in exercise or sleep studies. However, the RH measurement has not been validated. This paper has three parts. Part 1: In evaluating the sensor surface for RH detection, four Hygrochrons were placed on a wet paper towel. Two were placed on the towel with the protruding surface facing up and the other two facing down. The results showed that the Hygrochron with the protruding side was the sensor surface for detecting RH. Part 2: Twenty-seven Hygrochrons were calibrated in a humidity calibration chamber at a RH range from 40 to 90% at a constant temperature from 32 to 37°C. The mean bias was -1.08% between the Hygrochrons and the calibration chamber. The Hygrochron overestimated RH at the lower range (40-60%) and underestimated RH at the higher range (80-90%). The application of individual regression equations to each Hygrochron improved accuracy and reduced the mean bias to -0.002%. However, one Hygrochron showed outlier values that may be due to a manufacturing defect. Part 3: The reproducibility of Hygrochron for RH measurements was tested twice at the same condition of 35°C over a three-month interval. The intra-class coefficient was 0.996 to 1.000 with non-significant differences in the mean RH between test and re-test results (p=0.159). Hygrochrons are valid for RH measurements which show high reproducibility. It is recommended that Hygrochrons be calibrated over a range of desired RH and temperature prior to use to improve accuracy and detect any manufacturing defects.
在热舒适性研究中需要精确测量相对湿度(RH)。当皮肤附近温度或相对湿度超出热舒适区时,就会出现热不适。Hygrochron是一种可同时测量温度和相对湿度的小型无线设备,在运动或睡眠研究中既合适又方便。然而,其相对湿度测量尚未得到验证。本文分为三个部分。第1部分:在评估用于检测相对湿度的传感器表面时,将四个Hygrochron放在湿纸巾上。两个以突出面朝上放置在纸巾上,另外两个突出面朝下放置。结果表明,突出面的Hygrochron是用于检测相对湿度的传感器表面。第2部分:在湿度校准室中,于32至37°C的恒定温度下,在40%至90%的相对湿度范围内对27个Hygrochron进行校准。Hygrochron与校准室之间的平均偏差为-1.08%。Hygrochron在较低范围(40 - 60%)高估了相对湿度,而在较高范围(80 - 90%)低估了相对湿度。对每个Hygrochron应用单独的回归方程提高了准确性,并将平均偏差降低至-0.002%。然而,有一个Hygrochron显示出异常值,这可能是由于制造缺陷所致。第3部分:在35°C的相同条件下,间隔三个月对Hygrochron进行两次相对湿度测量的再现性测试。组内系数为0.996至1.000,测试和重新测试结果之间的平均相对湿度无显著差异(p = 0.159)。Hygrochron对于相对湿度测量是有效的,具有很高的再现性。建议在使用前在所需的相对湿度和温度范围内对Hygrochron进行校准,以提高准确性并检测任何制造缺陷。