IEEE Trans Biomed Eng. 2018 Apr;65(4):857-865. doi: 10.1109/TBME.2017.2723801. Epub 2017 Jul 5.
Mirror movements (MM) occur during unilateral actions and manifest as involuntary muscle activity of the passive limb, "mirroring" voluntary actions executed by the contralateral homologous body part. They are a normal motor feature in young children that gradually disappears. In children suffering from neurological disorders, e.g., unilateral cerebral palsy, MMs have been proposed to yield relevant information for diagnosis and therapy. However, in clinical practice, MM are typically assessed using an ordinal rating scale. Here, we introduce the grip force tracking (GriFT) device, a portable system to quantitatively assess MM during repetitive unimanual squeezing while playing a computer game.
The GriFT device consists of two handles, each equipped with two compressive force sensors (range 0-23 kg, F 1000 Hz). Children complete three trials of unimanual squeezing, whereby the visual display on the screen determines the squeezing rhythm (0.67 Hz at 15% maximum voluntary contraction, force-level adjusted per hand). MMs are characterized based on frequency, amplitude, and temporal features (synchronization, timing).
MM differed significantly between children with different clinical MM scores. MM frequency and amplitude were most discriminative. Categorization of physiological MM proved highly sensitive (89%-97%).
We demonstrated feasibility and validity of the GriFT device in a large cohort of typically developing children (N = 174, age 5-15 years), and its clinical applicability in children with unilateral cerebral palsy with various levels of hand function.
The quantification of MM is a promising tool to further investigate and categorize MM in children with unilateral cerebral palsy.
镜像运动(MM)发生在单侧运动中,表现为对侧同源肢体的被动肢体的不自主肌肉活动,“镜像”了对侧执行的主动运动。它们是儿童的正常运动特征,会逐渐消失。在患有神经障碍的儿童中,例如单侧脑瘫,MM 被认为可以提供有关诊断和治疗的相关信息。然而,在临床实践中,MM 通常使用序数评分量表进行评估。在这里,我们介绍了握力跟踪(GriFT)装置,这是一种便携式系统,可以在玩电脑游戏时进行重复单侧挤压时定量评估 MM。
GriFT 装置由两个手柄组成,每个手柄都配备有两个压缩力传感器(范围 0-23 千克,F 为 1000 Hz)。儿童完成三次单侧挤压试验,屏幕上的视觉显示确定挤压节奏(最大自主收缩的 15%,力水平根据每只手进行调整,频率为 0.67 Hz)。根据频率、幅度和时间特征(同步、定时)对 MM 进行特征描述。
具有不同临床 MM 评分的儿童之间的 MM 存在显著差异。MM 的频率和幅度最具区分性。对生理性 MM 的分类具有很高的敏感性(89%-97%)。
我们在一个典型的发展儿童队列(N = 174,年龄 5-15 岁)中证明了 GriFT 装置的可行性和有效性,并且在具有不同手部功能的单侧脑瘫儿童中具有临床适用性。
MM 的量化是进一步研究和分类单侧脑瘫儿童 MM 的有前途的工具。