A new era in gated myocardial perfusion imaging: Feasibility of data-driven cardiac contraction gating with multiple pinhole CZT SPECT.

机构信息

EA 7334 REMES, Université Paris-Diderot, Sorbonne Paris-Cité, Paris, France.

Nuclear Medicine Department, Cochin University Hospital, AP-HP, 27 rue du Faubourg Saint-Jacques, 75679, Paris Cedex 14, France.

出版信息

J Nucl Cardiol. 2018 Feb;25(1):257-268. doi: 10.1007/s12350-017-1010-0. Epub 2017 Aug 3.

Abstract

BACKGROUND

We previously validated the use of a data-driven cardiac respiratory-motion (RM) correction method (REGAT) applicable to CZT SPECT myocardial perfusion imaging (MPI). In this study, we adapted the same process used with REGAT for RM to generate data-driven cardiac contraction triggers and corresponding cardiac contraction-gated SPECT studies (GSPECT-DD). We aimed to study its feasibility and compare its performances to GSPECT studies generated with ECG monitor-based triggers (GSPECT-ECG).

METHODS

We included seven non-consecutive randomly chosen patients addressed for 1-day Tc-Tetrofosmin stress/rest MPI acquired with multi-pinhole CZT SPECT. We studied the degree of agreement between GSPECT-DD and GSPECT-ECG for the classification of acquired images into the 16 categories of mean cardiac cycle, and compared between the two methods the cine image quality and global LV systolic function of reconstructed studies.

RESULTS

We found almost perfect agreement between cardiac contraction triggers generated with data-driven and ECG monitor-based methods. As compared to GSPECT-ECG, GSPECT-DD provided comparable and well-correlated LV global systolic function parameters and similar cine image quality at both stress and rest.

CONCLUSIONS

Data-driven cardiac contraction gating using REGAT is feasible with low-dose and high-dose MPI CZT SPECT. It provides GSPECT-DD studies comparable to GSPECT-ECG.

摘要

背景

我们之前验证了一种适用于 CZT SPECT 心肌灌注成像(MPI)的基于数据的心脏呼吸运动(RM)校正方法(REGAT)的有效性。在这项研究中,我们采用了与 REGAT 相同的 RM 处理过程,生成了基于数据的心脏收缩触发,并相应地进行了心脏收缩门控 SPECT 研究(GSPECT-DD)。我们旨在研究其可行性,并将其性能与基于心电图监测器的触发生成的 GSPECT 研究(GSPECT-ECG)进行比较。

方法

我们纳入了 7 名非连续随机选择的患者,这些患者接受了 1 天 Tc-Tetrofosmin 应激/静息 MPI 多针孔 CZT SPECT 检查。我们研究了基于数据的和基于心电图监测器的心脏收缩触发在将采集的图像分类为 16 个平均心动周期类别方面的一致性,并比较了两种方法的电影图像质量和重建研究的整体左心室收缩功能。

结果

我们发现基于数据和基于心电图监测器的心脏收缩触发之间存在几乎完美的一致性。与 GSPECT-ECG 相比,GSPECT-DD 提供了可比较且相关性良好的 LV 整体收缩功能参数,并且在应激和静息时的电影图像质量相似。

结论

使用 REGAT 的基于数据的心脏收缩门控在低剂量和高剂量 MPI CZT SPECT 中是可行的。它提供了与 GSPECT-ECG 相当的 GSPECT-DD 研究。