J Nucl Cardiol. 2018 Feb;25(1):249-256. doi: 10.1007/s12350-016-0615-z. Epub 2016 Sep 27.
Gamma-cameras, with Cadmium-Zinc-Telluride (CZT) detectors, allow to perform myocardial perfusion imaging (MPI) with limited injected activities and recorded times. This study aimed at determining whether the routine assessment of left ventricular (LV) function with such limited counts protocols compares well with reference values from cardiac MRI.
The study included patients who have undergone cardiac MRI and an MPI routinely planned on a CZT camera with a low-dose protocol (120 MBq of Sestamibi for stress and 360 MBq at rest for 75 kg body weight), while targeting the recording of only 500 myocardial kcounts in order to limit the recording times (<10 minutes for stress, <4 minutes for rest). SPECT images were reconstructed with a method maintaining rather high spatial (8 mm) and temporal (16 frames/cycle) resolutions.
Seventy-six patients were included, and mean effective dose was 3.5 ± 1.7 mSv for the total MPI protocol. Correlations between CZT-SPECT and MRI were good to excellent for ejection fraction (r = 0.77), end-diastolic (r = 0.88) and end-systolic (r = 0.93) volumes, and the analysis of segmental contractility correlated well between the two techniques (kappa score = 0.72 ± 0.02).
LV function, assessed on a CZT camera with low injected activities and limited recording times, correlates well with the reference assessment from cardiac MRI.
伽马相机与碲锌镉(CZT)探测器结合使用,可在注射活动量有限和记录时间有限的情况下进行心肌灌注成像(MPI)。本研究旨在确定使用这种有限计数方案常规评估左心室(LV)功能是否与心脏 MRI 的参考值具有良好的可比性。
该研究纳入了在 CZT 相机上进行心脏 MRI 和常规 MPI 检查的患者,该相机采用低剂量方案(对于 75kg 体重,应激时注射 120MBq 甲氧基异丁基异腈,休息时注射 360MBq),目标是仅记录 500 个心肌计数以限制记录时间(应激时<10 分钟,休息时<4 分钟)。SPECT 图像使用一种方法进行重建,该方法保持较高的空间(8mm)和时间(16 帧/循环)分辨率。
共纳入 76 例患者,MPI 总方案的平均有效剂量为 3.5±1.7mSv。CZT-SPECT 与 MRI 之间的相关性对于射血分数(r=0.77)、舒张末期(r=0.88)和收缩末期(r=0.93)容积均为良好到极好,两种技术之间的节段收缩性分析也具有良好的相关性(kappa 评分=0.72±0.02)。
在注射活动量有限和记录时间有限的 CZT 相机上评估 LV 功能与心脏 MRI 的参考评估具有良好的相关性。