Suppr超能文献

即时检测血糖仪在细菌性脑膜炎中的应用。

Rapid detection of bacterial meningitis using a point-of-care glucometer.

机构信息

Emergency Department.

François Rabelais University, Tours, France.

出版信息

Eur J Emerg Med. 2019 Feb;26(1):41-46. doi: 10.1097/MEJ.0000000000000495.

Abstract

BACKGROUND

In case of acute bacterial meningitis, a decision on the need for intensive care admission should be made within the first hour. The aim of this study was to assess the ability of a point-of-care glucometer to determine abnormal cerebrospinal fluid (CSF) glucose concentration at the bedside that contributes toward bacterial meningitis diagnosis.

METHODS

We carried out a prospective study and simultaneously measured the glucose concentrations in CSF and blood using a central laboratory and a point-of-care glucometer. We compared CSF/blood glucose ratios obtained at the bedside with a glucometer versus those obtained by the central laboratory. We determined the performance characteristics of the CSF/blood glucose ratio provided by a glucometer to detect bacterial infection in the CSF immediately after CSF sampling.

RESULTS

We screened 201 CSF collection procedures during the study period and included 172 samples for analysis. Acute bacterial meningitis was diagnosed in 17/172 (9.9%) of CSF samples. The median turnaround time for a point-of-care glucometer analysis was 5 (interquartile range 2-10) min versus 112 (interquartile range 86-154) min for the central laboratory (P<0.0001). The optimal cut-off of the CSF/blood glucose ratio calculated from a bedside glucometer was 0.46, with a sensitivity of 94.1% (95% confidence interval: 71.3-99.9%), a specificity of 91% (95% confidence interval: 85.3-95%), and a positive likelihood ratio of 10.

CONCLUSION

A glucometer accurately detects an abnormal CSF/blood glucose ratio immediately after the lumbar puncture. This cheap point-of-care method has the potential to speed up the diagnostic process of patients with bacterial meningitis.

摘要

背景

在急性细菌性脑膜炎的情况下,应在最初的 1 小时内决定是否需要入住重症监护病房。本研究的目的是评估床边即时血糖仪测定脑脊液(CSF)葡萄糖浓度异常对诊断细菌性脑膜炎的作用。

方法

我们进行了一项前瞻性研究,同时使用中心实验室和床边即时血糖仪测量了 CSF 和血液中的葡萄糖浓度。我们比较了床边即时血糖仪与中心实验室获得的 CSF/血糖比值。我们确定了床边即时血糖仪检测 CSF 中细菌感染的 CSF/血糖比值的性能特征。

结果

研究期间共筛选了 201 次 CSF 采集过程,纳入 172 份样本进行分析。172 份 CSF 样本中诊断出急性细菌性脑膜炎 17/172(9.9%)。床边即时血糖仪分析的中位周转时间为 5(四分位间距 2-10)分钟,而中心实验室为 112(四分位间距 86-154)分钟(P<0.0001)。床边即时血糖仪计算的 CSF/血糖比值的最佳截断值为 0.46,灵敏度为 94.1%(95%置信区间:71.3-99.9%),特异性为 91%(95%置信区间:85.3-95%),阳性似然比为 10。

结论

血糖仪可准确检测到腰椎穿刺后立即出现的异常 CSF/血糖比值。这种廉价的床边即时检测方法有可能加快细菌性脑膜炎患者的诊断过程。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验