Measurements of cerebrospinal fluid (CSF) lactate can aid in detecting infections of the central nervous system and surrounding structures. Neurosurgical patients with temporary lumbar or ventricular CSF drainage harbor an increased risk for developing infections of the central nervous system, which require immediate therapeutic responses. Since blood gas analyzers enable rapid blood-lactate measurements, we were interested in finding out if we can reliably measure CSF-lactate by this point-of-care technique.
Neurosurgical patients on our intensive care unit (ICU) with either lumbar or external ventricular drainage due to a variety of reasons were included in this prospective observational study. Standard of care included measurements of leucocyte counts, total protein and lactate measurements in CSF by the neurochemical laboratory of our University Medical Center twice a week. With respect to this study, we additionally performed nearly daily measurements of cerebrospinal fluid by blood gas analyzers to determine the reliability of CSF-lactate measured by blood gas analyzers as compared to the standard measurements with a certified device.
62 patients were included in this study. We performed 514 CSF-lactate measurements with blood gas analyzers and compared 180 of these to the in-house standard CSF-lactate measurements. Both techniques correlated highly significantly (Pearson correlation index 0.94) even though lacking full concordance in a Bland-Altman plotting. Of particular importance, regular measurements enabled immediate detection of central infection in three patients who had developed meningitis during the course of their treatment.
Blood gas analyzers measure CSF-lactate with sufficient reliability and can help in the timely detection of a developing meningitis. In addition to and triggering established CSF diagnostics, CSF-lactate measurements by blood gas analyzers may improve surveillance of patients with CSF drainage. This study was retrospectively registered on April 20th 2020 in the German trial register. The trial registration number is DRKS00021466.
脑脊液(CSF)乳酸盐测量有助于检测中枢神经系统及周围结构的感染。接受临时腰椎或脑室CSF引流的神经外科患者发生中枢神经系统感染的风险增加,而这种感染需要立即进行治疗。由于血气分析仪能够快速测量血液乳酸盐,我们想了解是否可以通过这种即时检测技术可靠地测量CSF乳酸盐。
本前瞻性观察研究纳入了因各种原因在我们重症监护病房(ICU)接受腰椎或外部脑室引流的神经外科患者。常规护理包括我们大学医学中心神经化学实验室每周两次测量CSF中的白细胞计数、总蛋白和乳酸盐。关于本研究,我们还使用血气分析仪几乎每天测量脑脊液,以确定与使用经认证设备进行的标准测量相比,通过血气分析仪测量CSF乳酸盐的可靠性。
本研究纳入了62例患者。我们使用血气分析仪进行了514次CSF乳酸盐测量,并将其中180次与内部标准CSF乳酸盐测量结果进行了比较。尽管在Bland-Altman图中缺乏完全一致性,但两种技术的相关性非常显著(Pearson相关指数为0.94)。特别重要的是,定期测量能够立即检测出三名在治疗过程中发生脑膜炎的患者的中枢感染。
血气分析仪测量CSF乳酸盐具有足够的可靠性,有助于及时检测正在发生的脑膜炎。除了触发既定的CSF诊断外,通过血气分析仪测量CSF乳酸盐可能会改善对CSF引流患者的监测。本研究于2020年4月20日在德国试验注册中心进行了回顾性注册。试验注册号为DRKS00021466。