Al-Sheikh Mayss, Falavarjani Khalil Ghasemi, Pfau Maximilian, Uji Akihito, Le Phillip P, Sadda SriniVas R
Ophthalmic Surg Lasers Imaging Retina. 2017 Aug 1;48(8):623-631. doi: 10.3928/23258160-20170802-04.
To quantify vessel density (VD) and grey value (GV) as a measure of flow in the choriocapillaris (CC) in healthy subjects with optical coherence tomography angiography (OCTA).
In this prospective, noncomparative case series, 3 mm × 3 mm OCTA images of 36 eyes of 22 healthy individuals were obtained using a swept-source instrument. VD and GV levels were calculated on CC en face slabs in the central 1-mm (subfoveal field) and surrounding 2.5-mm parafoveal ring. VD was calculated as a ratio of vessel area over nonvessel area following image binarization. GV was computed as the mean, un-normalized greyscale intensity value for all pixels in the region of interest. For each eye, the procedure was repeated 1 minute to 2 minutes later and intersession repeatability was analyzed. The choroidal thickness (CT) was automatically measured in the subfoveal and parafoveal regions and compared to VD and GV values.
The VD ratio and GV was lower in the subfoveal field than in the parafoveal four sectors. The intersession intraclass correlation coefficients were high for both VD and GV measurements. There was no correlation observed between CT and VD or GV.
Quantitative metrics can be obtained from CC OCTA en face images. These values show moderate to good intersession repeatability. These normative data may be of value as a reference of comparison in future studies of eyes with disease. [Ophthalmic Surg Lasers Imaging Retina. 2017;48:623-631.].
利用光学相干断层扫描血管造影(OCTA)对健康受试者脉络膜毛细血管(CC)中的血管密度(VD)和灰度值(GV)进行量化,以衡量血流情况。
在这个前瞻性、非对照病例系列研究中,使用扫频源仪器获取了22名健康个体36只眼的3mm×3mm OCTA图像。在中央1mm(黄斑中心凹下区域)和周围2.5mm的黄斑旁环的CC表面平板上计算VD和GV水平。VD通过图像二值化后血管面积与非血管面积的比值来计算。GV计算为感兴趣区域内所有像素的平均灰度强度值(未标准化)。对每只眼睛,在1至2分钟后重复该操作,并分析不同测量时段之间的可重复性。在黄斑中心凹下和黄斑旁区域自动测量脉络膜厚度(CT),并与VD和GV值进行比较。
黄斑中心凹下区域的VD比值和GV低于黄斑旁的四个象限。VD和GV测量的不同测量时段组内相关系数均较高。未观察到CT与VD或GV之间存在相关性。
可从CC的OCTA表面图像中获得定量指标。这些值显示出中度到良好的不同测量时段之间的可重复性。这些标准化数据可能作为未来眼病研究中比较的参考具有价值。[《眼科手术、激光与视网膜成像》。2017;48:623 - 631。]