Repeatability, interocular correlation and agreement of quantitative swept-source optical coherence tomography angiography macular metrics in healthy subjects.

机构信息

Joint Shantou International Eye Center, Shantou University & the Chinese University of Hong Kong, Shantou, China.

The Chinese University of Hong Kong, Hong Kong, China.

出版信息

Br J Ophthalmol. 2019 Mar;103(3):415-420. doi: 10.1136/bjophthalmol-2018-311874. Epub 2018 May 29.

Abstract

PURPOSE

To investigate the repeatability, interocular correlation and agreement of quantitative swept-source optical coherence tomography angiography (SS-OCTA) metrics in healthy subjects.

METHODS

Thirty-three healthy normal subjects were enrolled. The macula was scanned four times by an SS-OCTA system using the 3 mm×3 mm mode. The superficial capillary map images were analysed using a MATLAB program. A series of parameters were measured: foveal avascular zone (FAZ) area, FAZ perimeter, FAZ circularity, parafoveal vessel density, fractal dimension and vessel diameter index (VDI). The repeatability of four scans was determined by intraclass correlation coefficient (ICC). Then the averaged results were analysed for intereye difference, correlation and agreement using paired t-test, Pearson's correlation coefficient (r), ICC and Bland-Altman plot.

RESULTS

The repeatability assessment of the macular metrics exported high ICC values (ranged from 0.853 to 0.996). There is no statistically significant difference in the OCTA metrics between the two eyes. FAZ area (ICC=0.961, r=0.929) and FAZ perimeter (ICC=0.884, r=0.802) showed excellent binocular correlation. Fractal dimension (ICC=0.732, r=0.578) and VDI (ICC=0.707, r=0.547) showed moderate binocular correlation, while parafoveal vessel density had poor binocular correlation. Bland-Altman plots showed the range of agreement was from -0.0763 to 0.0954 mm for FAZ area and from -0.0491 to 0.1136 for parafoveal vessel density.

CONCLUSIONS

The macular metrics obtained using SS-OCTA showed excellent repeatability in healthy subjects. We showed high intereye correlation in FAZ area and perimeter, moderate correlation in fractal dimension and VDI, while vessel density had poor correlation in normal healthy subjects.

摘要

目的

研究健康受试者定量扫频光相干断层扫描血管造影(SS-OCTA)测量值的重复性、眼间相关性和一致性。

方法

纳入 33 名健康正常受试者。使用 SS-OCTA 系统以 3mm×3mm 模式对黄斑进行四次扫描。使用 MATLAB 程序分析浅层毛细血管图图像。测量一系列参数:中心凹无血管区(FAZ)面积、FAZ 周长、FAZ 圆度、旁中心凹血管密度、分形维数和血管直径指数(VDI)。采用组内相关系数(ICC)评估四次扫描的重复性。然后使用配对 t 检验、Pearson 相关系数(r)、ICC 和 Bland-Altman 图分析双眼间的平均结果差异、相关性和一致性。

结果

黄斑测量值的重复性评估得出的 ICC 值较高(范围为 0.853 至 0.996)。双眼 OCTA 测量值之间无统计学差异。FAZ 面积(ICC=0.961,r=0.929)和 FAZ 周长(ICC=0.884,r=0.802)显示出极好的双眼相关性。分形维数(ICC=0.732,r=0.578)和 VDI(ICC=0.707,r=0.547)显示出中等的双眼相关性,而旁中心凹血管密度的双眼相关性较差。Bland-Altman 图显示 FAZ 面积的一致性范围为-0.0763 至 0.0954mm,旁中心凹血管密度的一致性范围为-0.0491 至 0.1136mm。

结论

SS-OCTA 获得的黄斑测量值在健康受试者中表现出极好的重复性。我们在 FAZ 面积和周长方面显示出高度的眼间相关性,在分形维数和 VDI 方面显示出中等的相关性,而在正常健康受试者中,血管密度的相关性较差。