University of California, San Francisco, 505 Parnassus Ave, San Francisco, CA, 94143, USA.
Med Phys. 2017 Nov;44(11):5718-5725. doi: 10.1002/mp.12525. Epub 2017 Sep 22.
To determine whether axial or helical mode is more appropriate for a 16 cm collimation CT scanner capable of step-and-shoot volumetric axial coverage, in terms of radiation dose, image quality, and scan duration.
All scans were performed with a Revolution CT (GE Healthcare) operating at 120 kV and 100 mAs. Using calibrated optically stimulated luminescence detectors, radiation dose along the axial scan profile was evaluated at the isocenter, including the overlap region between two axial sections. This overlap region measures 3 cm in the z-axis at the isocenter and is required to obtain sufficient projection data from the relatively large cone-beam angles. Using an image quality phantom (Gammex Model 464), spatial resolution, CT number uniformity, image noise, and low contrast detectability (LCD) were evaluated under five different conditions: in the middle of a helical acquisition, in the middle of a 16 cm axial section, at both ends of an axial section and in the overlap region between two axial sections. Scan durations and dose length products (DLP) were recorded for prescribed scan lengths of 2.5-100 cm.
The overlap region between two axial sections received a dose 83% higher than the single-exposure region at the isocenter. Within a single axial section, the dose at the anode end was 37% less than at the cathode end due to the anode heel effect. Image noise ranged from a low of 13 HU for the cathode end of an axial section up to 14.7 HU for the anode end (P < 0.001). The LCD was at lower at the anode end of the axial section compared to both the cathode end (P < 0.05) and the overlap location (P < 0.02). The spatial resolution and CT number uniformity were consistent among all conditions. Scan durations were shorter (0.28 s) for the axial mode compared to the helical mode at scan lengths ≤ 16 cm, and longer at scan lengths ≥ 16 cm where more than one table position was required, up to a difference of 13.9 s for a the 100 cm scan length (3.8 s for helical compared to 17.6 s for axial). DLPs were consistent between scan modes; slightly lower in axial mode at shorter scan lengths due to helical overranging, and slightly higher in axial mode at longer scan lengths due to the axial overlap regions.
To ensure the most consistent radiation dose and image quality along the scan length, we recommend helical mode for scans longer than the 16 cm coverage of a single axial section. For scan lengths ≤ 16 cm, axial scanning is the most practical option, with a shorter scan duration and higher dose efficiency.
在辐射剂量、图像质量和扫描时间方面,确定对于能够进行步进式容积轴向覆盖的 16cm 准直 CT 扫描仪,轴向或螺旋模式哪种更合适。
所有扫描均在 Revolution CT(通用电气医疗)上进行,管电压为 120kV,管电流为 100mA。使用经过校准的光激励发光探测器,在等中心评估轴向扫描轮廓线上的辐射剂量,包括两个轴向切片之间的重叠区域。在等中心,该重叠区域在 z 轴上为 3cm,需要从相对较大的锥形束角度获得足够的投影数据。使用图像质量体模(Gammex Model 464),在五种不同条件下评估空间分辨率、CT 数均匀性、图像噪声和低对比度检测能力(LCD):在螺旋采集的中间、16cm 轴向切片的中间、轴向切片的两端以及两个轴向切片之间的重叠区域。记录规定的扫描长度为 2.5-100cm 时的扫描时间和剂量长度乘积(DLP)。
两个轴向切片之间的重叠区域在等中心处比单次曝光区域接收的剂量高 83%。在单个轴向切片内,由于阳极脚跟效应,阳极端的剂量比阴极端低 37%。由于阳极脚跟效应,阳极端的剂量比阴极端低 37%。图像噪声范围从轴向切片阴极端的低 13HU 到阳极端的 14.7HU(P<0.001)。与轴向切片的阴极端(P<0.05)和重叠位置(P<0.02)相比,轴向切片的阳极端的 LCD 较低。在所有条件下,空间分辨率和 CT 数均匀性一致。在扫描长度≤16cm 时,轴向模式的扫描时间(0.28s)比螺旋模式短,而在需要多个台位的扫描长度≥16cm 时,扫描时间更长,最长相差 13.9s,即 100cm 扫描长度时(3.8s 用于螺旋,17.6s 用于轴向)。扫描模式之间的 DLP 一致;在较短的扫描长度下,轴向模式稍低,这是由于螺旋扫描的超范围;在较长的扫描长度下,轴向模式稍高,这是由于轴向重叠区域。
为了确保沿扫描长度获得最一致的辐射剂量和图像质量,我们建议对于超过单个轴向切片 16cm 覆盖范围的扫描,使用螺旋模式。在扫描长度≤16cm 时,轴向扫描是最实用的选择,扫描时间更短,剂量效率更高。