一次性(“康查帕克”)和可重复使用的加湿系统对呼吸机相关性肺炎发病率的影响。

Influence of disposable ('Conchapak') and reusable humidifying systems on the incidence of ventilation pneumonia.

作者信息

Daschner F, Kappstein I, Schuster F, Scholz R, Bauer E, Joossens D, Just H

机构信息

Department of Hospital Epidemiology, University Hospital of Freiburg, FRG.

出版信息

J Hosp Infect. 1988 Feb;11(2):161-8. doi: 10.1016/0195-6701(88)90059-x.

Abstract

The contamination of disposable ('Conchapak') and reusable humidifying systems and their influence on the incidence of pneumonia was studied in 116 patients requiring continuous mechanical ventilation therapy. The water reservoirs of 11 (15.9%) of the 69 disposable systems became colonized, but all reusable systems were found to be sterile. In four of the 11 samples, the organisms isolated corresponded with those cultured from tracheal secretions several days before. Ventilator-associated pneumonia occurred in 36 (31.0%) of the patients, but there was no statistically significant difference in the incidence of pneumonia between the patients treated with the disposable or the reusable humidifying systems. Gram-negative bacteria were the predominant organisms isolated from tracheal aspirates of patients who developed ventilator-associated pneumonia. These results suggest that disposable humidifying systems do not influence the rate of ventilator-associated pneumonia in mechanically ventilated patients.

摘要

在116例需要持续机械通气治疗的患者中,研究了一次性(“Conchapak”)和可重复使用的加湿系统的污染情况及其对肺炎发病率的影响。69个一次性系统中的11个(15.9%)的储水槽被定植,但所有可重复使用的系统均被发现无菌。在11个样本中的4个中,分离出的微生物与几天前从气管分泌物中培养出的微生物一致。36例(31.0%)患者发生了呼吸机相关性肺炎,但使用一次性或可重复使用的加湿系统治疗的患者之间肺炎发病率无统计学显著差异。革兰氏阴性菌是从发生呼吸机相关性肺炎的患者气管吸出物中分离出的主要微生物。这些结果表明,一次性加湿系统不会影响机械通气患者呼吸机相关性肺炎的发生率。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验