Suppr超能文献

苹果手表™测量最大心率的有效性及设备间差异

The validity and inter-device variability of the Apple Watch™ for measuring maximal heart rate.

作者信息

Abt Grant, Bray James, Benson Amanda Clare

机构信息

a School of Life Sciences , The University of Hull , Kingston upon Hull , UK.

b Department of Health and Medical Sciences , Swinburne University of Technology , Melbourne , Australia.

出版信息

J Sports Sci. 2018 Jul;36(13):1447-1452. doi: 10.1080/02640414.2017.1397282. Epub 2017 Nov 1.

Abstract

Maximal heart rate (HR) is a fundamental measure used in exercise prescription. The Apple Watch™ measures heart rate yet the validity and inter-device variability of the device for measuring HR are unknown. Fifteen participants completed a maximal oxygen uptake test while wearing an Apple Watch™ on each wrist. Criterion HR was measured using a Polar T31™ chest strap. There were good to very good correlations between the watches and criterion (left: r = 0.87 [90%CI: 0.67 to 0.95]; right: r = 0.98 [90%CI: 0.94 to 0.99]). Standardised mean bias for the left and right watches compared to the criterion were 0.14 (90%CI: -0.12 to 0.39; trivial) and 0.04 (90%CI: -0.07 to 0.15; trivial). Standardised typical error of the estimate for the left and right watches compared to the criterion were 0.51 (90%CI: 0.38 to 0.80; moderate) and 0.22 (90%CI: 0.16 to 0.34; small). Inter-device standardised typical error was 0.46 (90%CI: 0.36 to 0.68; moderate), ICC = 0.84 (90%CI: 0.65 to 0.93). The Apple Watch™ has good to very good criterion validity for measuring HR, with no substantial under- or over-estimation. There were moderate and small prediction errors for the left and right watches. Inter-device variability in HR is moderate.

摘要

最大心率(HR)是运动处方中使用的一项基本指标。苹果手表™可测量心率,但该设备测量心率的有效性及设备间的变异性尚不清楚。15名参与者在每只手腕上佩戴一块苹果手表™的同时完成了最大摄氧量测试。使用 Polar T31™ 胸带测量标准心率。手表测量值与标准值之间的相关性良好至极佳(左手:r = 0.87 [90%CI:0.67至0.95];右手:r = 0.98 [90%CI:0.94至0.99])。与标准值相比,左手和右手手表的标准化平均偏差分别为0.14(90%CI:-0.12至0.39;微不足道)和0.04(90%CI:-0.07至0.15;微不足道)。与标准值相比,左手和右手手表的标准化估计典型误差分别为0.51(90%CI:0.38至0.80;中等)和0.22(90%CI:0.16至0.34;较小)。设备间标准化典型误差为0.46(90%CI:0.36至0.68;中等),组内相关系数(ICC)= 0.84(90%CI:0.65至0.93)。苹果手表™在测量心率方面具有良好至极佳的标准效度,不存在明显的低估或高估情况。左手和右手手表存在中等和较小的预测误差。心率的设备间变异性为中等。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索