利用智能手环评估体育课中学生的心率:可行性、可靠性和有效性研究。

Using Smart Bracelets to Assess Heart Rate Among Students During Physical Education Lessons: Feasibility, Reliability, and Validity Study.

机构信息

School of Physical Education and Sport Training, Shanghai University of Sport, Shanghai, China.

Shanghai Research Center for Physical Fitness and Health of Children and Adolescents, Shanghai University of Sport, Shanghai, China.

出版信息

JMIR Mhealth Uhealth. 2020 Aug 5;8(8):e17699. doi: 10.2196/17699.

Abstract

BACKGROUND

An increasing number of wrist-worn wearables are being examined in the context of health care. However, studies of their use during physical education (PE) lessons remain scarce.

OBJECTIVE

We aim to examine the reliability and validity of the Fizzo Smart Bracelet (Fizzo) in measuring heart rate (HR) in the laboratory and during PE lessons.

METHODS

In Study 1, 11 healthy subjects (median age 22.0 years, IQR 3.75 years) twice completed a test that involved running on a treadmill at 6 km/h for 12 minutes and 12 km/h for 5 minutes. During the test, participants wore two Fizzo devices, one each on their left and right wrists, to measure their HR. At the same time, the Polar Team2 Pro (Polar), which is worn on the chest, was used as the standard. In Study 2, we went to 10 schools and measured the HR of 24 students (median age 14.0 years, IQR 2.0 years) during PE lessons. During the PE lessons, each student wore a Polar device on their chest and a Fizzo on their right wrist to measure HR data. At the end of the PE lessons, the students and their teachers completed a questionnaire where they assessed the feasibility of Fizzo. The measurements taken by the left wrist Fizzo and the right wrist Fizzo were compared to estimate reliability, while the Fizzo measurements were compared to the Polar measurements to estimate validity. To measure reliability, intraclass correlation coefficients (ICC), mean difference (MD), standard error of measurement (SEM), and mean absolute percentage errors (MAPE) were used. To measure validity, ICC, limits of agreement (LOA), and MAPE were calculated and Bland-Altman plots were constructed. Percentage values were used to estimate the feasibility of Fizzo.

RESULTS

The Fizzo showed excellent reliability and validity in the laboratory and moderate validity in a PE lesson setting. In Study 1, reliability was excellent (ICC>0.97; MD<0.7; SEM<0.56; MAPE<1.45%). The validity as determined by comparing the left wrist Fizzo and right wrist Fizzo was excellent (ICC>0.98; MAPE<1.85%). Bland-Altman plots showed a strong correlation between left wrist Fizzo measurements (bias=0.48, LOA=-3.94 to 4.89 beats per minute) and right wrist Fizzo measurements (bias=0.56, LOA=-4.60 to 5.72 beats per minute). In Study 2, the validity of the Fizzo was lower compared to that found in Study 1 but still moderate (ICC>0.70; MAPE<9.0%). The Fizzo showed broader LOA in the Bland-Altman plots during the PE lessons (bias=-2.60, LOA=-38.89 to 33.69 beats per minute). Most participants considered the Fizzo very comfortable and easy to put on. All teachers thought the Fizzo was helpful.

CONCLUSIONS

When participants ran on a treadmill in the laboratory, both left and right wrist Fizzo measurements were accurate. The validity of the Fizzo was lower in PE lessons but still reached a moderate level. The Fizzo is feasible for use during PE lessons.

摘要

背景

越来越多的腕戴可穿戴设备在医疗保健领域得到了研究。然而,关于它们在体育课(PE)中的使用的研究仍然很少。

目的

我们旨在研究 Fizzo Smart Bracelet(Fizzo)在实验室和体育课中测量心率(HR)的可靠性和有效性。

方法

在研究 1 中,11 名健康受试者(中位数年龄 22.0 岁,IQR 3.75 岁)两次完成了一项测试,包括在跑步机上以 6 公里/小时跑 12 分钟和 12 公里/小时跑 5 分钟。在测试过程中,参与者佩戴两个 Fizzo 设备,一个在左腕,一个在右腕,以测量他们的 HR。同时,使用佩戴在胸部的 Polar Team2 Pro(Polar)作为标准。在研究 2 中,我们前往 10 所学校,在体育课上测量 24 名学生(中位数年龄 14.0 岁,IQR 2.0 岁)的 HR。在体育课上,每个学生在胸部佩戴一个 Polar 设备,在右腕佩戴一个 Fizzo 来测量 HR 数据。在体育课结束时,学生和他们的老师完成了一份问卷,评估 Fizzo 的可行性。比较左手腕 Fizzo 和右手腕 Fizzo 的测量值来估计可靠性,同时将 Fizzo 的测量值与 Polar 的测量值进行比较来估计有效性。为了测量可靠性,使用了组内相关系数(ICC)、平均差异(MD)、测量误差的标准误差(SEM)和平均绝对百分比误差(MAPE)。为了测量有效性,计算了 ICC、限值协议(LOA)和 MAPE,并绘制了 Bland-Altman 图。百分比值用于估计 Fizzo 的可行性。

结果

Fizzo 在实验室和体育课中均表现出优异的可靠性和有效性,在体育课中表现出中等有效性。在研究 1 中,可靠性非常好(ICC>0.97;MD<0.7;SEM<0.56;MAPE<1.45%)。通过比较左手腕 Fizzo 和右手腕 Fizzo 得出的有效性非常好(ICC>0.98;MAPE<1.85%)。Bland-Altman 图显示左手腕 Fizzo 测量值(偏差=0.48,LOA=-3.94 至 4.89 次/分钟)和右手腕 Fizzo 测量值(偏差=0.56,LOA=-4.60 至 5.72 次/分钟)之间存在很强的相关性。在研究 2 中,与研究 1 相比,Fizzo 的有效性较低,但仍处于中等水平(ICC>0.70;MAPE<9.0%)。在体育课期间,Fizzo 的 Bland-Altman 图中的 LOA 更宽(偏差=-2.60,LOA=-38.89 至 33.69 次/分钟)。大多数参与者认为 Fizzo 非常舒适且易于佩戴。所有老师都认为 Fizzo 很有帮助。

结论

当参与者在实验室的跑步机上跑步时,左右腕 Fizzo 的测量值都是准确的。在体育课中,Fizzo 的有效性较低,但仍达到中等水平。Fizzo 可用于体育课。