Preload, Coefficient of Friction, and Thread Friction in an Implant-Abutment-Screw Complex.

作者信息

Wentaschek Stefan, Tomalla Sven, Schmidtmann Irene, Lehmann Karl Martin

出版信息

Int J Prosthodont. 2017 Nov/Dec;30(6):542-544. doi: 10.11607/ijp.5286.

Abstract

PURPOSE

To examine the screw preload, coefficient of friction (COF), and tightening torque needed to overcome the thread friction of an implant-abutment-screw complex.

MATERIALS AND METHODS

In a customized load frame, 25 new implant-abutment-screw complexes including uncoated titanium alloy screws were torqued and untorqued 10 times each, applying 25 Ncm.

RESULTS

Mean preload values decreased significantly from 209.8 N to 129.5 N according to the number of repetitions. The overall COF increased correspondingly. There was no comparable trend for the thread friction component.

CONCLUSION

These results suggest that the application of a used implant-abutment-screw complex may be unfavorable for obtaining optimal screw preload.

摘要

目的

研究种植体-基台-螺钉复合体克服螺纹摩擦力所需的螺钉预紧力、摩擦系数(COF)和拧紧扭矩。

材料与方法

在定制的加载框架中,对25个包括未涂层钛合金螺钉的新型种植体-基台-螺钉复合体进行扭矩施加和松开,各进行10次,施加扭矩为25 Ncm。

结果

根据重复次数,平均预紧力值从209.8 N显著降至129.5 N。总体摩擦系数相应增加。螺纹摩擦分量没有类似趋势。

结论

这些结果表明,使用过的种植体-基台-螺钉复合体可能不利于获得最佳螺钉预紧力。