Zipprich Holger, Rathe Florian, Pinz Sören, Schlotmann Luca, Lauer Hans-Christoph, Ratka Christoph
Int J Oral Maxillofac Implants. 2018 Mar/Apr;33(2):e25-e32. doi: 10.11607/jomi.5837.
The aim of this study was to investigate the effects of tightening torque, screw head angle, and thread number on the preload force of abutment screws.
The test specimens consisted of three self-manufactured components (ie, a thread sleeve serving as an implant analog, an abutment analog, and an abutment screw). The abutment screws were fabricated with metric M1.6 external threads. The thread number varied between one and seven threads. The screw head angles were produced in eight varying angles (30 to 180 degrees). A sensor unit simultaneously measured the preload force of the screw and the torsion moment inside the screw shank. The tightening of the screw with the torque wrench was performed in five steps (15 to 35 Ncm). The torque wrench was calibrated before each step.
Only the tightening torque and screw head angle affected the resulting preload force of the implant-abutment connection. The thread number had no effect. There was an approximately linear correlation between tightening torque and preload force.
The tightening torque and screw head angle were the only study parameters that affected the resulting preload force of the abutment screw. The results obtained from this experiment are valid only for a single torque condition. Further investigations are needed that analyze other parameters that affect preload force. Once these parameters are known, it will add value for a strong, but detachable connection between the implant and abutment. Short implants and flat-to-flat connections especially will benefit significantly from this knowledge.
本研究旨在探讨拧紧扭矩、螺钉头角度和螺纹数量对基台螺钉预紧力的影响。
测试样本由三个自制部件组成(即一个用作种植体模拟物的螺纹套筒、一个基台模拟物和一个基台螺钉)。基台螺钉采用公制M1.6外螺纹制造。螺纹数量在1至7个螺纹之间变化。螺钉头角度设置为八个不同角度(30至180度)。一个传感器单元同时测量螺钉的预紧力和螺钉杆内部的扭矩。使用扭矩扳手分五步(15至35 Ncm)拧紧螺钉。在每一步之前对扭矩扳手进行校准。
仅拧紧扭矩和螺钉头角度会影响种植体 - 基台连接的最终预紧力。螺纹数量没有影响。拧紧扭矩和预紧力之间存在近似线性关系。
拧紧扭矩和螺钉头角度是仅有的影响基台螺钉最终预紧力的研究参数。本实验获得的结果仅适用于单一扭矩条件。需要进一步研究分析其他影响预紧力的参数。一旦了解这些参数,将有助于实现种植体与基台之间牢固但可拆卸的连接。特别是短种植体和平面对平面连接将从这一知识中显著受益。