Development of a Blood Pressure Measurement Instrument with Active Cuff Pressure Control Schemes.

机构信息

Department of Electrical Engineering, National Taiwan University of Science and Technology, Taipei, Taiwan.

Department of Biomedical Engineering, National Defense Medical Center, Taipei, Taiwan.

出版信息

J Healthc Eng. 2017;2017:9128745. doi: 10.1155/2017/9128745. Epub 2017 Oct 3.

Abstract

This paper presents an oscillometric blood pressure (BP) measurement approach based on the active control schemes of cuff pressure. Compared with conventional electronic BP instruments, the novelty of the proposed BP measurement approach is to utilize a variable volume chamber which actively and stably alters the cuff pressure during inflating or deflating cycles. The variable volume chamber is operated with a closed-loop pressure control scheme, and it is activated by controlling the piston position of a single-acting cylinder driven by a screw motor. Therefore, the variable volume chamber could significantly eliminate the air turbulence disturbance during the air injection stage when compared to an air pump mechanism. Furthermore, the proposed active BP measurement approach is capable of measuring BP characteristics, including systolic blood pressure (SBP) and diastolic blood pressure (DBP), during the inflating cycle. Two modes of air injection measurement (AIM) and accurate dual-way measurement (ADM) were proposed. According to the healthy subject experiment results, AIM reduced 34.21% and ADM reduced 15.78% of the measurement time when compared to a commercial BP monitor. Furthermore, the ADM performed much consistently (i.e., less standard deviation) in the measurements when compared to a commercial BP monitor.

摘要

本文提出了一种基于袖带压力主动控制方案的示波法血压(BP)测量方法。与传统的电子血压计相比,所提出的 BP 测量方法的新颖之处在于利用了一个可变容积室,该容积室在充气或放气循环期间主动且稳定地改变袖带压力。可变容积室采用闭环压力控制方案进行操作,并通过控制由螺杆电机驱动的单作用缸的活塞位置来激活。因此,与气泵机构相比,可变容积室在空气注入阶段可以显著消除空气湍流干扰。此外,所提出的主动 BP 测量方法能够在充气循环期间测量包括收缩压(SBP)和舒张压(DBP)在内的 BP 特征。提出了两种空气注入测量模式(AIM)和精确双向测量(ADM)。根据健康受试者的实验结果,与商用血压计相比,AIM 减少了 34.21%的测量时间,ADM 减少了 15.78%的测量时间。此外,与商用血压计相比,ADM 在测量中的表现更为一致(即标准差更小)。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/7f23/5651164/9cf06e32fb42/JHE2017-9128745.001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验