Blood pressure measurement using finger cuff.

作者信息

Lee J, Choi E, Jeong H, Kim K, Park J

机构信息

Samsung Advanced Institute of Technology, Mt. 14-1, Nongseo-Ri, Giheung-Eup, Yongin-Si, Gyeonggi-Do, Republic of Korea 449-712 (phone: 81-31-280-6525; fax: 82-31-280-9208; e-mail:

出版信息

Conf Proc IEEE Eng Med Biol Soc. 2005;2005:3575-7. doi: 10.1109/IEMBS.2005.1617252.

Abstract

Many research groups have studied blood pressure measurement in finger artery because of its convenience. But, low accuracy prohibits many hypertension patients from using this device. So, we suggest measurement algorithm that measure systolic and diastolic blood pressure in finger artery. And we also develop calibration method that decreases the error from difference of finger circumference by subjects. We apply our methods for 90 subjects (age form 20 to 49, 55 male, 35 female) to test feasibility of our method by AAMI SP10 standard. The mean difference of our system is ±4.7mmHg for systolic pressure, ±4.2mmHg for systolic pressure. It proved that the feasibility of our method is clinically acceptable.(under ±5mmHg).