Yu Sha-Sha, Song Hui, Tang Xin
Clinical College of Ophthalmology, Tianjin Medical University, Tianjin 300020, China.
Tianjin Eye Hospital, Tianjin 300020, China.
Int J Ophthalmol. 2017 Nov 18;10(11):1703-1709. doi: 10.18240/ijo.2017.11.11. eCollection 2017.
To determine the repeatability of Ophtha Top topography and assess the consistency with intraocular lens (IOL)-Master and LenstarLS900 (Lenstar) in measuring corneal parameters among cataract patients.
Totally 125 eyes were enrolled. Corneas were successively measured with Ophtha Top, IOL-Master and Lenstar at least three times. The flattest meridian power (Kf), the steepest meridian power (Ks), mean power (Km), J0 and J45 were recorded. Intra-class correlation coefficients (ICCs), the coefficient of variance (COV), within subject standard deviation (Sw), and test-retest repeatability (2.77Sw) were adopted to determine the repeatability. The 95% limit of agreement (95%LOA) and Bland-Altman plots were used to assess comparability.
Repeatability of Ophtha Top topography for measuring corneal parameters showed the ICCs were all above 0.93, 2.77Sw was lower than 0.31, and the COV of the Kf and Ks was lower than 0.25. The keratometric readings with Ophtha Top topography were flatter than with the IOL-Master and Lenstar devices, while the Pearson correlation coefficients were over 0.97. The J0 and J45 with Ophtha Top topography were smaller compared with Lenstar and IOL-Master, while was comparable between Lenstar and IOL-Master.
Ophtha Top topography shows excellent repeatability for measuring corneal parameters. However, differences between the Ophtha TOP topography and Lenstar, IOL-Master both in cornea curvature and the astigmatism should be noted clinically.
确定Ophtha Top角膜地形图的可重复性,并评估其在白内障患者角膜参数测量中与眼内晶状体(IOL)Master及Lenstar LS900(Lenstar)的一致性。
共纳入125只眼。使用Ophtha Top、IOL-Master和Lenstar对角膜进行至少三次连续测量。记录最平坦子午线屈光力(Kf)、最陡峭子午线屈光力(Ks)、平均屈光力(Km)、J0和J45。采用组内相关系数(ICC)、变异系数(COV)、受试者内标准差(Sw)以及重测可重复性(2.77Sw)来确定可重复性。使用95%一致性界限(95%LOA)和Bland-Altman图评估可比性。
Ophtha Top角膜地形图测量角膜参数的可重复性显示,ICC均高于0.93,2.77Sw低于0.31,Kf和Ks的COV低于0.25。Ophtha Top角膜地形图测量的角膜曲率读数比IOL-Master和Lenstar设备测得的更平坦,而Pearson相关系数超过0.97。与Lenstar和IOL-Master相比,Ophtha Top角膜地形图测量的J0和J45较小,而Lenstar和IOL-Master之间则具有可比性。
Ophtha Top角膜地形图在测量角膜参数方面显示出优异的可重复性。然而,临床上应注意Ophtha Top角膜地形图与Lenstar、IOL-Master在角膜曲率和散光方面的差异。