Abstract

AIM

To determine the repeatability of Ophtha Top topography and assess the consistency with intraocular lens (IOL)-Master and LenstarLS900 (Lenstar) in measuring corneal parameters among cataract patients.

METHODS

Totally 125 eyes were enrolled. Corneas were successively measured with Ophtha Top, IOL-Master and Lenstar at least three times. The flattest meridian power (Kf), the steepest meridian power (Ks), mean power (Km), J0 and J45 were recorded. Intra-class correlation coefficients (ICCs), the coefficient of variance (COV), within subject standard deviation (Sw), and test-retest repeatability (2.77Sw) were adopted to determine the repeatability. The 95% limit of agreement (95%LOA) and Bland-Altman plots were used to assess comparability.

RESULTS

Repeatability of Ophtha Top topography for measuring corneal parameters showed the ICCs were all above 0.93, 2.77Sw was lower than 0.31, and the COV of the Kf and Ks was lower than 0.25. The keratometric readings with Ophtha Top topography were flatter than with the IOL-Master and Lenstar devices, while the Pearson correlation coefficients were over 0.97. The J0 and J45 with Ophtha Top topography were smaller compared with Lenstar and IOL-Master, while was comparable between Lenstar and IOL-Master.

CONCLUSION

Ophtha Top topography shows excellent repeatability for measuring corneal parameters. However, differences between the Ophtha TOP topography and Lenstar, IOL-Master both in cornea curvature and the astigmatism should be noted clinically.

摘要

目的

确定Ophtha Top角膜地形图的可重复性,并评估其在白内障患者角膜参数测量中与眼内晶状体(IOL)Master及Lenstar LS900(Lenstar)的一致性。

方法

共纳入125只眼。使用Ophtha Top、IOL-Master和Lenstar对角膜进行至少三次连续测量。记录最平坦子午线屈光力(Kf)、最陡峭子午线屈光力(Ks)、平均屈光力(Km)、J0和J45。采用组内相关系数(ICC)、变异系数(COV)、受试者内标准差(Sw)以及重测可重复性(2.77Sw)来确定可重复性。使用95%一致性界限(95%LOA)和Bland-Altman图评估可比性。

结果

Ophtha Top角膜地形图测量角膜参数的可重复性显示,ICC均高于0.93,2.77Sw低于0.31,Kf和Ks的COV低于0.25。Ophtha Top角膜地形图测量的角膜曲率读数比IOL-Master和Lenstar设备测得的更平坦,而Pearson相关系数超过0.97。与Lenstar和IOL-Master相比,Ophtha Top角膜地形图测量的J0和J45较小,而Lenstar和IOL-Master之间则具有可比性。

结论

Ophtha Top角膜地形图在测量角膜参数方面显示出优异的可重复性。然而,临床上应注意Ophtha Top角膜地形图与Lenstar、IOL-Master在角膜曲率和散光方面的差异。

相似文献

1
Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.
Int J Ophthalmol. 2017 Nov 18;10(11):1703-1709. doi: 10.18240/ijo.2017.11.11. eCollection 2017.
2
Evaluation of the repeatability of the Lenstar and comparison with two other non-contact biometric devices in myopes.
Clin Exp Optom. 2013 Jan;96(1):92-9. doi: 10.1111/j.1444-0938.2012.00793.x. Epub 2012 Sep 3.
3
Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace.
PLoS One. 2016 Jan 11;11(1):e0147086. doi: 10.1371/journal.pone.0147086. eCollection 2016.
5
Repeatability and agreement of ocular biometry measurements: Aladdin versus Lenstar.
Br J Ophthalmol. 2017 Sep;101(9):1223-1229. doi: 10.1136/bjophthalmol-2016-309365. Epub 2017 Jan 27.
10
Comparison of corneal power, corneal astigmatism, and axis location in normal eyes obtained from an autokeratometer and a corneal topographer.
J Cataract Refract Surg. 2012 Apr;38(4):648-54. doi: 10.1016/j.jcrs.2011.11.026. Epub 2012 Feb 18.

引用本文的文献

1
Comparison of keratometry data using handheld and table-mounted instruments in healthy adults.
Int Ophthalmol. 2021 Oct;41(10):3451-3458. doi: 10.1007/s10792-021-01909-8. Epub 2021 Jun 9.

本文引用的文献

1
Accuracy of Different Topographic Instruments in Calculating Corneal Power after Myopic Photorefractive Keratectomy.
J Ophthalmic Vis Res. 2017 Jul-Sep;12(3):254-259. doi: 10.4103/jovr.jovr_74_16.
2
Comparison of corneal power obtained from VERION image-guided surgery system and four other devices.
Clin Ophthalmol. 2017 Jul 12;11:1291-1299. doi: 10.2147/OPTH.S137878. eCollection 2017.
3
Evaluation of biometry and corneal astigmatism in cataract surgery patients from Central China.
BMC Ophthalmol. 2017 Apr 26;17(1):56. doi: 10.1186/s12886-017-0450-2.
4
Prevalence of corneal astigmatism before cataract surgery in Indian population.
Int Ophthalmol. 2017 Jun;37(3):683-689. doi: 10.1007/s10792-016-0327-z. Epub 2016 Aug 27.
6
Age-related changes of the ocular surface: a hospital setting-based retrospective study.
J Ophthalmol. 2014;2014:532378. doi: 10.1155/2014/532378. Epub 2014 Aug 11.
7
Prevalence of Corneal Astigmatism in Patients before Cataract Surgery in Northern China.
J Ophthalmol. 2014;2014:536412. doi: 10.1155/2014/536412. Epub 2014 Jun 3.
8
The effect of head tilt on keratometric measurement using the IOLMaster.
Eye (Lond). 2013 Dec;27(12):1411-7. doi: 10.1038/eye.2013.209. Epub 2013 Sep 20.
9
Agreement between a partial coherence interferometer and 2 manual keratometers.
J Cataract Refract Surg. 2013 Oct;39(10):1550-60. doi: 10.1016/j.jcrs.2013.03.034. Epub 2013 Jul 19.
10
Repeatability, reproducibility, and agreement of corneal power measurements obtained with a new corneal topographer.
J Cataract Refract Surg. 2013 Oct;39(10):1561-9. doi: 10.1016/j.jcrs.2013.04.029. Epub 2013 Jul 13.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验