Int Ophthalmol. 2024 Feb 20;44(1):103. doi: 10.1007/s10792-024-03046-4.
This study aimed to evaluate the consistency of preoperative keratometric values, anterior segment, and intraocular lens (IOL) power measurements in patients with cataract and no comorbidities using the Sirius topography device (CSO, Italy) and Lenstar LS 900 (Haag-Streit AG, Köeniz, Switzerland).
Patients with grade 2 and 3 cataracts who applied to Ophthalmology Clinic of Dışkapı Yıldırım Beyazıt Education and Research Hospital, University of Health Sciences and planned for cataract surgery were included the study. Forty eyes with cataract from 40 patients were taken in the study. All patients underwent preoperative assessment using a combined Scheimpflug-Placido disc-based tomography device (Sirius) and Lenstar before cataract surgery. Keratometric measurements, such as flat keratometry (K1), steep keratometry (K2), and maximum keratometry (Kmax), and anterior segment parameters, white-to-white (WTW) distance, IOL power, astigmatism (AST), anterior chamber depth (ACD), aqueous depth (AD), and central cornea thickness (CCT), were recorded.
There were significant differences between K1 and K1, K2 and K2, Kmax and Kmax, WTW and WTW, and IOL power versus IOL power. However, there were insignificant differences between AST and AST, ACD versus ACD, AD and AD, and CCT and CCT variables. Furthermore, it was found that Sirius measured significantly higher than Lenstar, especially in terms of IOL power.
Significant differences were observed between Lenstar and Sirius in terms of keratometric values, WTW distance, and IOL power. The IOL power value measured with Sirius was found to be higher than the IOL power value measured with Lenstar.
本研究旨在评估使用 Sirius 角膜地形图仪(CSO,意大利)和 Lenstar LS 900(Haag-Streit AG,Köniz,瑞士)对无合并症的白内障患者进行术前角膜曲率值、眼前节和人工晶状体(IOL)测量的一致性。
本研究纳入了向健康科学大学 Dışkapı Yıldırım Beyazıt 教育与研究医院眼科诊所就诊并计划行白内障手术的 2 级和 3 级白内障患者。本研究纳入了 40 例 40 只眼的白内障患者。所有患者在白内障手术前均使用联合 Scheimpflug-Placido 盘式基于断层的角膜地形图仪(Sirius)和 Lenstar 进行术前评估。记录角膜曲率测量值,如平角膜曲率(K1)、陡角膜曲率(K2)和最大角膜曲率(Kmax),以及眼前节参数,如角膜直径(WTW)、IOL 度数、散光(AST)、前房深度(ACD)、房水深度(AD)和中央角膜厚度(CCT)。
K1 与 K1、K2 与 K2、Kmax 与 Kmax、WTW 与 WTW 和 IOL 度数与 IOL 度数之间存在显著差异。然而,AST 与 AST、ACD 与 ACD、AD 与 AD 和 CCT 与 CCT 变量之间无显著差异。此外,发现 Sirius 测量的 IOL 度数明显高于 Lenstar,尤其是在 IOL 度数方面。
Lenstar 和 Sirius 在角膜曲率值、WTW 距离和 IOL 度数方面存在显著差异。使用 Sirius 测量的 IOL 度数高于使用 Lenstar 测量的 IOL 度数。