Penna Rachele R, de Sanctis Ugo, Catalano Martina, Brusasco Luca, Grignolo Federico M
Department of Surgical Sciences, Eye Clinic, University of Torino, Via Juvarra 19, Torino 10122, Italy.
Int J Ophthalmol. 2017 Mar 18;10(3):453-460. doi: 10.18240/ijo.2017.03.20. eCollection 2017.
To compare the repeatability/reproducibility of measurement by high-resolution Placido disk-based topography with that of a high-resolution rotating Scheimpflug camera and assess the agreement between the two instruments in measuring corneal power in eyes with keratoconus and post-laser keratomileusis (LASIK).
One eye each of 36 keratoconic patients and 20 subjects who had undergone LASIK was included in this prospective observational study. Two independent examiners worked in a random order to take three measurements of each eye with both instruments. Four parameters were measured on the anterior cornea: steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), and astigmatism (Ks-Kf). Intra-examiner repeatability and inter-examiner reproducibility were evaluated by calculating the within-subject standard deviation (Sw) the coefficient of repeatability (R), the coefficient of variation (CoV), and the intraclass correlation coefficient (ICC). Agreement between instruments was tested with the Bland-Altman method by calculating the 95% limits of agreement (95% LoA).
In keratoconic eyes, the intra-examiner and inter-examiner ICC were >0.95. As compared with measurement by high-resolution Placido disk-based topography, the intra-examiner R of the high-resolution rotating Scheimpflug camera was lower for Kf (0.32 0.88), Ks (0.61 0.88), and Km (0.32 0.84) but higher for Ks-Kf (0.70 0.57). Inter-examiner R values were lower for all parameters measured using the high-resolution rotating Scheimpflug camera. The 95% LoA were -1.28 to +0.55 for Kf, -1.36 to +0.99 for Ks, -1.08 to +0.50 for Km, and -1.11 to +1.48 for Ks-Kf. In the post-LASIK eyes, the intra-examiner and inter-examiner ICC were >0.87 for all parameters. The intra-examiner and inter-examiner R were lower for all parameters measured using the high-resolution rotating Scheimpflug camera. The intra-examiner R was 0.17 0.88 for Kf, 0.21 0.88 for Ks, 0.17 0.86 for Km, and 0.28 0.33 for Ks-Kf. The inter-examiner R was 0.09 0.64 for Kf, 0.15 0.56 for Ks, 0.09 0.59 for Km, and 0.18 0.23 for Ks-Kf. The 95% LoA were -0.54 to +0.58 for Kf, -0.51 to +0.53 for Ks and Km, and -0.28 to +0.27 for Ks-Kf.
As compared with Placido disk-based topography, the high-resolution rotating Scheimpflug camera provides more repeatable and reproducible measurements of Ks, Kf and Ks in keratoconic and post-LASIK eyes. Agreement between instruments is fair in keratoconus and very good in post-LASIK eyes.
比较基于高分辨率普拉西多盘的角膜地形图测量与高分辨率旋转式Scheimpflug相机测量的重复性/再现性,并评估这两种仪器在测量圆锥角膜和准分子激光原位角膜磨镶术(LASIK)术后眼睛角膜屈光力方面的一致性。
这项前瞻性观察研究纳入了36例圆锥角膜患者和20例接受过LASIK手术患者的各一只眼睛。两名独立的检查者按随机顺序使用两种仪器对每只眼睛进行三次测量。在前角膜上测量四个参数:陡峭角膜曲率(Ks)、平坦角膜曲率(Kf)、平均角膜曲率(Km)和散光(Ks-Kf)。通过计算受试者内标准差(Sw)、重复性系数(R)、变异系数(CoV)和组内相关系数(ICC)来评估检查者内重复性和检查者间再现性。使用Bland-Altman方法通过计算95%一致性界限(95% LoA)来测试仪器之间的一致性。
在圆锥角膜眼中,检查者内和检查者间的ICC均>0.95。与基于高分辨率普拉西多盘的角膜地形图测量相比,高分辨率旋转式Scheimpflug相机在Kf(0.32对0.88)、Ks(0.61对0.88)和Km(0.32对0.84)的检查者内R较低,但在Ks-Kf(0.70对0.57)方面较高。使用高分辨率旋转式Scheimpflug相机测量的所有参数的检查者间R值均较低。Kf的95% LoA为-1.28至+0.55,Ks为-1.36至+0.99,Km为-1.08至+0.50,Ks-Kf为-1.11至+1.48。在LASIK术后眼中,所有参数的检查者内和检查者间ICC均>0.87。使用高分辨率旋转式Scheimpflug相机测量的所有参数的检查者内和检查者间R均较低。Kf的检查者内R为0.17对0.88,Ks为0.21对0.88,Km为0.17对0.86,Ks-Kf为0.28对0.33。Kf的检查者间R为0.09对0.64,Ks为0.15对0.56,Km为0.09对0.59,Ks-Kf为0.18对0.23。Kf的95% LoA为-0.54至+0.58,Ks和Km为-0.51至+0.53,Ks-Kf为-0.28至+0.27。
与基于普拉西多盘的角膜地形图相比,高分辨率旋转式Scheimpflug相机在圆锥角膜和LASIK术后眼中对Ks、Kf和Km的测量具有更高的重复性和再现性。两种仪器在圆锥角膜中的一致性尚可,在LASIK术后眼中非常好。